Skip to content

CERTAIN® GOLD-TITE® IUNIHGRecalls & reported problems

ZimVie US Corp LLC · CERTAIN® GOLD-TITE® HEXED SCREW · Class 2 · Implant

Quick summary

This ZimVie Certain Gold-Tite hexed screw is used to secure dental abutments on implants. Its 152 adverse event reports include 7 injuries and no deaths, and many describe screws that loosen, fracture or fail to seat fully, causing crowns and abutments to shift or requiring removal. In the past 12 months there were 12 reports, down from 37 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
151
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 151 adverse event reports that list this exact device (DI 00844868008538), received between Jun 18, 2020 and May 19, 2026 — classified as 7 injuries, 144 malfunctions.

The most frequently reported specific device problem is “Unintended Movement” (81 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 6 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 3 reportsApr 2024: 1 reportsMay 2024: 4 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 7 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 10 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 4 reportsSep 2025: 1 reportsOct 2025: 4 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 3 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Movement81
  • Fracture67
  • Adverse Event Without Identified Device or Use Problem2
  • Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions121
  • No Information13
  • Insufficient Information5
  • Inflammation4
  • Pain4
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Discomfort1
  • Failure of Implant1

Recent adverse event reports

The 25 most recent of 151 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2026Unintended Movement

    IT WAS REPORTED THAT THE GOLD SCREW WAS NOT ABLE TO SEAT ALL THE WAY. THE SCREW WAS LOOSE AND THUS THE WHOLE CROWN/ABUTMENT HAD MOVEMENT. WE WERE ABLE TO RETRIEVE THE GOLD SCREW, BUT THE ABUTMENT WAS LODGED INTO THE IMPLANT AND WOULD NOT COME OUT. THE ABUTMENT AND CROWN HAD TO BE CUT TO BE REMOVED. TOOTH #13. THE IMPLANT IS OK AND I STILL HAVE THE REMAINING CROWN AND ABUTMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25216153 ↗
  • MalfunctionMay 4, 2026Fracture

    IT WAS REPORTED A FRACTURE OF THREE OF THE PROSTHETIC SCREWS AT THE THREADED PART IN A PATIENT WITH A MANDIBULAR HYBRID PROSTHESIS ON FOUR IMPLANTS (TOOTH SITE 32,34,42,44), SO IT WAS DECIDED TO REPLACE ALL OF THEM, TAKING INTO ACCOUNT THE POSSIBLE OCLUSAL OVERLOAD. PATIENT SUFFERED PROSTHESIS MOBILITY AND INFLAMMATION.

    Patient problems: Inflammation

    MAUDE report 25064950 ↗
  • MalfunctionApr 21, 2026Fracture

    IT WAS REPORTED THAT THERE WAS A SCREW FRACTURED INSIDE OF THE IMPLANT AT TOOTH SITE #21. THE SCREW WAS SUCTIONED DURING REMOVAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24941729 ↗
  • MalfunctionMar 30, 2026Unintended Movement

    IT WAS REPORTED THAT EITHER THE SCREW CAME LOOSE OR DROPPED WHEN SEATING THE ABUTMENT, BUT THE PATIENT MAY HAVE SWALLOWED IT. (IT DISAPPEARED).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24714152 ↗
  • MalfunctionMar 24, 2026Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24682267 ↗
  • MalfunctionMar 5, 2026Fracture

    IT WAS REPORTED THAT CLINICIAN CALLED AND REPORTED SCREW FRACTURED AN WAS REMOVED FROM IMPLANT. PATIENT HAS NO IMPACT JUST NEEDS REPLACEMENT SCREW. THE COMPLAINT PRODUCT WAS DISCARDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24524403 ↗
  • MalfunctionFeb 20, 2026Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24399322 ↗
  • MalfunctionJan 9, 2026Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24018704 ↗
  • MalfunctionDec 11, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23778878 ↗
  • MalfunctionDec 5, 2025Unintended Movement; Unintended Movement

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Discomfort; Pain

    MAUDE report 23726379 ↗
  • MalfunctionDec 4, 2025Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23715877 ↗
  • MalfunctionOct 15, 2025Fracture; Fracture

    IT WAS REPORTED THAT A FRACTURE PORTION WAS FOUND INSIDE THE IMPLANT, SCREW IUNIHG FRACTURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23302410 ↗
  • MalfunctionOct 15, 2025Unintended Movement; Unintended Movement

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23301448 ↗
  • MalfunctionOct 15, 2025Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT SCREW LOOSENED AND WAS RETIGHTENED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23300620 ↗
  • InjuryOct 15, 2025Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: Inflammation

    MAUDE report 23300264 ↗
  • MalfunctionSep 12, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23043146 ↗
  • MalfunctionAug 27, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22903212 ↗
  • MalfunctionAug 27, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22903108 ↗
  • MalfunctionAug 25, 2025Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22891287 ↗
  • MalfunctionAug 12, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE SCREW FRACTURED INSIDE OF THE IMPLANT. THE IMPLANT REMAINS IN THE PATIENT'S MOUTH WITHOUT ANY ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22784137 ↗
  • MalfunctionJul 1, 2025Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT THE SCREW LOOSENED AT TOOTH SITE #19 A FEW MONTHS AFTER THE ABUTMENT WAS SEATED IN THE PATIENT'S MOUTH AND THE SCREW WAS ULTIMATELY REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22374759 ↗
  • MalfunctionJul 1, 2025Unintended Movement; Unintended Movement

    IT WAS REPORTED A LOOSENING DETECTED DUE TO CROWN MOBILITY ON TOOTH SITE #25. THE CROWN WAS FRACTURED. PROCEDURE WAS COMPLETED CHANGING THE SCREW. NO IMPACT ON THE PATIENT REPORTED. PATIENT WILL RETURN TO PLACE THE NEW PROSTHESIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22368827 ↗
  • MalfunctionJun 25, 2025Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT THE SCREW LOOSENED IN THE PATIENT'S MOUTH. REMOVED AND REPLACED WITH ANOTHER SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22333525 ↗
  • MalfunctionMay 28, 2025Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22096642 ↗
  • MalfunctionMay 21, 2025Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT ABUTMENT CAME LOOSE WOULDN¿T SEAT ALL THE WAY AND PATIENT IN PATIENTS MOUTH, SWALLOWED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22065661 ↗

Other ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS devices

All ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS devices →