Nobel Biocare® 36662Recalls & reported problems
Nobel Biocare AB · Temporary Abutment Non-Engaging Conical Connection RP · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07332747034301), received between Aug 24, 2019 and Aug 24, 2019 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Component Misassembled” (5 reports).
2 FDA recall records name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Aug 9, 2019Class IITerminated
Packages of abutments may contain an incorrect screw. While the screw will tighten, it cannot be guaranteed that the screw will not loosen. In case the screw loosens the abutment should be retightened with the correct screw.
Z-1852-2020 · Root cause: Process control · Recall details · FDA record ↗
- Aug 9, 2019Class IITerminated
Packages of abutments may contain an incorrect screw. While the screw will tighten, it cannot be guaranteed that the screw will not loosen. In case the screw loosens the abutment should be retightened with the correct screw.
Z-1853-2020 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Component Misassembled5
Patient problems
- No Known Impact Or Consequence To Patient5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 24, 2019Component Misassembled; Component Misassembled
NOBEL BIOCARE RECEIVED THIS COMPLAINT ON (B)(6) 2019 FROM A CUSTOMER IN (B)(6) STATING THAT AN INCORRECT SCREW WAS PACKED WITH ARTICLE 36662 (TEMPORARY ABUTMENT NON-ENGAGING CC RP), LOT 13078005. THE INCORRECT SCREWS WERE NOT USED ON A PATIENT. INSPECTION OF THE RETAIN AND COMPLAINT SAMPLES CONFIRMED THAT THE BLISTERS CONTAINED HALB 20034 (CATALOG NUMBER 37893) INSTEAD OF THE INTENDED SCREW HALB 12320 (CATALOG NUMBER 37892). MANUFACTURER INVESTIGATION: INVESTIGATION SHOWED THAT ARTICLE 36662 WAS ASSEMBLED WITH HALB 12320 SCREWS FROM BATCH 12122536. IMMEDIATE ACTIONS TAKEN, E.G. QUALITY HOLD/BATCH RESTRICTION INCL. DATE: ARTICLE 36662, LOT 13078005 WAS RESTRICTED ON JUNE 11, 2019. FOLLOWING THE PRELIMINARY INVESTIGATION, LOTS 13077622, 13077550, 13076822, 13077560 AND 13077700 WERE RESTRICTED ON JUNE 27, 2019 AND LOT 13077551 WAS RESTRICTED ON JULY 10, 2019. ROOT CAUSE: THE ROOT CAUSE WILL BE FURTHER INVESTIGATED IN THE CAPA THAT WAS INITIATED (CAPA #(B)(4)).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8928660 ↗MalfunctionAug 24, 2019Component Misassembled; Component Misassembled
NOBEL BIOCARE RECEIVED THIS COMPLAINT ON (B)(6) 2019 FROM A CUSTOMER IN (B)(6) STATING THAT AN INCORRECT SCREW WAS PACKED WITH ARTICLE 36662 (TEMPORARY ABUTMENT NON-ENGAGING CC RP), LOT 13078005. THE INCORRECT SCREWS WERE NOT USED ON A PATIENT. INSPECTION OF THE RETAIN AND COMPLAINT SAMPLES CONFIRMED THAT THE BLISTERS CONTAINED HALB 20034 (CATALOG NUMBER 37893) INSTEAD OF THE INTENDED SCREW HALB 12320 (CATALOG NUMBER 37892). MANUFACTURER INVESTIGATION. INVESTIGATION SHOWED THAT ARTICLE 36662 WAS ASSEMBLED WITH HALB 12320 SCREWS FROM BATCH 12122536. IMMEDIATE ACTIONS TAKEN, E.G. QUALITY HOLD/BATCH RESTRICTION INCL. DATE: ARTICLE 36662, LOT 13078005 WAS RESTRICTED ON (B)(6) 2019. FOLLOWING THE PRELIMINARY INVESTIGATION, LOTS 13077622, 13077550, 13076822, 13077560 AND 13077700 WERE RESTRICTED ON (B)(6) 2019 AND LOT 13077551 WAS RESTRICTED ON (B)(6) 2019. ROOT CAUSE: THE ROOT CAUSE WILL BE FURTHER INVESTIGATED IN THE CAPA THAT WAS INITIATED (CAPA #(B)(4)).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8928659 ↗MalfunctionAug 24, 2019Component Misassembled; Component Misassembled
NOBEL BIOCARE RECEIVED THIS COMPLAINT ON MAY 23, 2019 FROM A CUSTOMER IN (B)(6) STATING THAT AN INCORRECT SCREW WAS PACKED WITH ARTICLE 36662 (TEMPORARY ABUTMENT NON-ENGAGING CC RP), LOT 13078005. THE INCORRECT SCREWS WERE NOT USED ON A PATIENT. INSPECTION OF THE RETAIN AND COMPLAINT SAMPLES CONFIRMED THAT THE BLISTERS CONTAINED HALB 20034 (CATALOG NUMBER 37893) INSTEAD OF THE INTENDED SCREW HALB 12320 (CATALOG NUMBER 37892). MANUFACTURER INVESTIGATION: INVESTIGATION SHOWED THAT ARTICLE 36662 WAS ASSEMBLED WITH HALB 12320 SCREWS FROM BATCH 12122536. IMMEDIATE ACTIONS TAKEN, E.G. QUALITY HOLD/BATCH RESTRICTION INCL. DATE: ARTICLE 36662, LOT 13078005 WAS RESTRICTED ON JUNE 11, 2019. FOLLOWING THE PRELIMINARY INVESTIGATION, LOTS 13077622, 13077550, 13076822, 13077560 AND 13077700 WERE RESTRICTED ON JUNE 27, 2019 AND LOT 13077551 WAS RESTRICTED ON JULY 10, 2019. ROOT CAUSE: THE ROOT CAUSE WILL BE FURTHER INVESTIGATED IN THE CAPA THAT WAS INITIATED (CAPA #(B)(4)).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8928658 ↗MalfunctionAug 24, 2019Component Misassembled; Component Misassembled
NOBEL BIOCARE RECEIVED THIS COMPLAINT ON (B)(6) 2019 FROM A CUSTOMER IN (B)(6) STATING THAT AN INCORRECT SCREW WAS PACKED WITH ARTICLE 36662 (TEMPORARY ABUTMENT NON-ENGAGING CC RP), LOT 13078005. THE INCORRECT SCREWS WERE NOT USED ON A PATIENT. INSPECTION OF THE RETAIN AND COMPLAINT SAMPLES CONFIRMED THAT THE BLISTERS CONTAINED HALB 20034 (CATALOG NUMBER 37893) INSTEAD OF THE INTENDED SCREW HALB 12320 (CATALOG NUMBER 37892). MANUFACTURER INVESTIGATION: INVESTIGATION SHOWED THAT ARTICLE 36662 WAS ASSEMBLED WITH HALB 12320 SCREWS FROM BATCH 12122536. IMMEDIATE ACTIONS TAKEN, E.G. QUALITY HOLD/BATCH RESTRICTION INCL. DATE: ARTICLE 36662, LOT 13078005 WAS RESTRICTED ON JUNE 11, 2019. FOLLOWING THE PRELIMINARY INVESTIGATION, LOTS 13077622, 13077550, 13076822, 13077560 AND 13077700 WERE RESTRICTED ON JUNE 27, 2019 AND LOT 13077551 WAS RESTRICTED ON JULY 10, 2019. ROOT CAUSE THE ROOT CAUSE WILL BE FURTHER INVESTIGATED IN THE CAPA THAT WAS INITIATED (CAPA #(B)(4)).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8928657 ↗MalfunctionAug 24, 2019Component Misassembled; Component Misassembled
NOBEL BIOCARE RECEIVED THIS COMPLAINT ON MAY 23, 2019 FROM A CUSTOMER IN (B)(6) STATING THAT AN INCORRECT SCREW WAS PACKED WITH ARTICLE 36662 (TEMPORARY ABUTMENT NON-ENGAGING CC RP), LOT 13078005. THE INCORRECT SCREWS WERE NOT USED ON A PATIENT. INSPECTION OF THE RETAIN AND COMPLAINT SAMPLES CONFIRMED THAT THE BLISTERS CONTAINED HALB 20034 (CATALOG NUMBER 37893) INSTEAD OF THE INTENDED SCREW HALB 12320 (CATALOG NUMBER 37892). MANUFACTURER INVESTIGATION: INVESTIGATION SHOWED THAT ARTICLE 36662 WAS ASSEMBLED WITH HALB 12320 SCREWS FROM BATCH 12122536. IMMEDIATE ACTIONS TAKEN, E.G. QUALITY HOLD/BATCH RESTRICTION INCL. DATE: ARTICLE 36662, LOT 13078005 WAS RESTRICTED ON JUNE 11, 2019. FOLLOWING THE PRELIMINARY INVESTIGATION, LOTS 13077622, 13077550, 13076822, 13077560 AND 13077700 WERE RESTRICTED ON JUNE 27, 2019 AND LOT 13077551 WAS RESTRICTED ON JULY 10, 2019. ROOT CAUSE THE ROOT CAUSE WILL BE FURTHER INVESTIGATED IN THE CAPA THAT WAS INITIATED (CAPA #(B)(4)).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8928656 ↗
Other ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS devices
- CERTAIN® GOLD-TITE® IUNIHGZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS151 reports
- CERTAIN® IUNIHTZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS67 reports
- ATLANTIS® 35502Dentsply Ih, Inc. · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS66 reports
- ATLANTIS® 35502Dentsply Ih, Inc. · Abutment, implant, dental, endosseous54 reports
- Esthetic Abutment 36669Nobel Biocare AB · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall2 reports
- CERTAIN® BELLATEK® IEDAT5ZimVie US Corp LLC · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall1 report
- Custom Abutments Final AbutmentsInstitut Straumann AG · Abutment, implant, dental, endosseous1 recall0 reports
- P.004 RC CARES TITANIUM AND CERAMIC ABUTMENTS Final AbutmentsInstitut Straumann AG · ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page