Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00845225001117), received between Nov 24, 2018 and Nov 24, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (1 report).

1 other recall cover devices cleared under the same FDA submission (K111395); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Separation1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 24, 2018Material Separation; Material Separation

    THIS CASE WAS REVIEWED AND INVESTIGATED ACCORDING TO THE MANUFACTURER'S POLICY. THE CUSTOMER REPORTED THAT DURING PROCEDURE ERROR CODE "ATTACH WIRE TO CONNECTOR" WAS RECEIVED. THE DEVICE WAS VISUALLY AND MICROSCOPICALLY INSPECTED AND DAMAGE WAS OBSERVED. THE PRESSURE SENSOR WAS BROKEN AND THE DISTAL PORTION WAS MISSING. THE WIRE WAS NOT RECOGNIZED DURING ZERO TESTING. THE PROBABLE CAUSE OF THE REPORTED FAILURE IS AN ELECTRICAL FAILURE WITH THE WIRE. THE DEVICE WAS OBSERVED TO HAVE A BROKEN SENSOR, WHICH IS CONSISTENT WITH THE REPORTED FAILURE. THE MANUFACTURING DOCUMENTATION FOR THIS DEVICE WAS REVIEWED AND THE DEVICE MET ALL QUALITY AND MANUFACTURING RELEASE CRITERIA. THIS PRODUCT PROBLEM IS BEING REPORTED BECAUSE A PORTION OF THE SENSOR IS MISSING FROM THE DEVICE, IT COULD NOT CONCLUSIVELY BE DETERMINED WHEN THE SEPARATION OCCURRED. THERE IS A POTENTIAL FOR HARM IF THE MALFUNCTION WERE TO RECUR.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8100314 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K111395. One submission often covers many models, so check the product and lot details in each record.

  • Z-1845-2015 · Jun 4, 2015 · Terminated
    Prestige Plus 185cm J-Tip: Part number: 9185J; Product Usage: The Prestige Plus Wire is indicated for use to measure pressure in blood vessels including both coronary and periphe

Other Wire, Guide, Catheter devices

All Wire, Guide, Catheter devices →