Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00845225003067), received between Oct 2, 2023 and Oct 2, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Entrapment of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Entrapment of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 2, 2023Entrapment of Device; Entrapment of Device

    IT WAS REPORTED THAT A RECONNAISSANCE CATHETER WAS USED IN A DIAGNOSTIC PERIPHERAL PROCEDURE IN A SEVERELY CALCIFIED DISTAL SFA. DURING INSERTION OF THE CATHETER THROUGH AN INTRODUCER SHEATH AND OVER A NON-PHILIPS GUIDEWIRE, THE GUIDEWIRE GOT CAUGHT AND UNABLE TO EXIT OUT OF THE CATHETER GUIDEWIRE EXIT PORT (Y-CONNECTOR). THE CATHETER WAS REMOVED, RE-FLUSHED, AND REINSERTED INTO THE PATIENT, BUT DID NOT RESOLVE. THEREFORE, THE PHYSICIAN CUT OFF A PIECE OF THE Y-CONNECTOR IN ORDER TO ALLOW GUIDEWIRE MOVEMENT. AFTER, THE CATHETER WAS ABLE TO ADVANCE FURTHER DOWN THE GUIDEWIRE TOWARD THE STENOSIS. THERE WAS NO PATIENT INJURY REPORTED. THIS ADVERSE EVENT AND PRODUCT PROBLEM IS BEING SUBMITTED BECAUSE A NON-PHILIPS GUIDEWIRE GOT STUCK INSIDE THE RECONNAISSANCE CATHETER, AND ADDITIONAL INTERVENTION WAS REQUIRED FOR GUIDEWIRE MOVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17852702 ↗