The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00845699000388), received between May 8, 2025 and Mar 9, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Electrical Overstress” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Electrical Overstress3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 9, 2026Electrical Overstress

    THE BLOWER WAS SENT TO AGILITI FOR REPAIR WITH THE REPORTED ISSUE OF POWER FAILURE. UPON OPENING THE DEVICE, AN AGILITI TECHNICIAN DISCOVERED INTERNAL BURN DAMAGE THAT CAUSED THE POWER FAILURE. BURN DAMAGE WAS OBSERVED ON THE SIDE PANEL ASSEMBLY, AND A MARK APPROXIMATELY 6 INCHES IN LENGTH WAS PRESENT ON THE INTERIOR SURFACE OF THE UNIT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24551961 ↗
  • MalfunctionDec 30, 2025Electrical Overstress; Electrical Overstress

    DEVICE WAS SENT IN FOR REPAIR AND UPON OPENING THE UNIT, THE TECHNICIAN FOUND BURN DAMAGE INTERNALLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23926672 ↗
  • MalfunctionMay 8, 2025Electrical Overstress; Electrical Overstress

    THE PUMP WAS SENT TO AGILITI'S CENTER OF EXCELLENCE (COE) FOR REPAIR OF A COMPONENT UNRELATED TO THIS ISSUE. DURING DEVICE INSPECTION, THE COE IDENTIFIED BURN MARKS INSIDE THE PUMP; NO EXTERIOR DAMAGE WAS FOUND. NO COMPLAINTS OR REPORTS REGARDING THIS ISSUE WERE FILED BY THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21985546 ↗