Skip to content

SUN PLATBRecalls & reported problems

Sizewise Rentals, L.L.C. · 120V Platinum 6000 · Class 1

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00845699003624), received between Jan 22, 2026 and Jan 22, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Electrical /Electronic Property Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 22, 2026Electrical /Electronic Property Problem

    THE PUMP WAS SENT TO AGILITI FOR REPAIR WITH THE REPORTED ISSUE THAT THE UNIT HAD NO POWER. UPON OPENING THE DEVICE, AN AGILITI TECHNICIAN DISCOVERED INTERNAL BURN DAMAGE THAT CAUSED THE POWER FAILURE. THE BURN MARK WAS LOCATED ON THE POWER BOARD AND MEASURED APPROXIMATELY 1 CM IN DIAMETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24146676 ↗
  • MalfunctionJan 22, 2026Electrical /Electronic Property Problem

    THE PUMP WAS SENT TO AGILITI FOR REPAIR WITH THE REPORTED ISSUE THAT THE UNIT HAD NO POWER WHEN PLUGGED IN AND WOULD NOT TURN ON. UPON OPENING THE DEVICE, AN AGILITI TECHNICIAN DISCOVERED INTERNAL BURN DAMAGE THAT CAUSED THE POWER FAILURE. THE BURN MARK WAS LOCATED ON THE POWER BOARD AND MEASURED APPROXIMATELY 5 MM IN DIAMETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24146428 ↗

Other Device, Transfer, Patient, Manual devices

All Device, Transfer, Patient, Manual devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.