Portable Overhead Lift LIFTOHRecalls & reported problems
Sizewise Rentals, L.L.C. · Portable Overhead Lift & Transfer · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jan 25, 2024Class IIOpen, Classified
Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.
Z-1595-2024 · Root cause: Device Design · Recall details · FDA record ↗
Other Lift, Patient, Ac-Powered devices
- Bedside Lift LIFTBDSizewise Rentals, L.L.C. · Lift, Patient, Ac-Powered1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page