Skip to content

Portable Overhead Lift LIFTOHRecalls & reported problems

Sizewise Rentals, L.L.C. · Portable Overhead Lift & Transfer · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jan 25, 2024Class IIOpen, Classified

    Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.

    Z-1595-2024 · Root cause: Device Design · Recall details · FDA record ↗

Other Lift, Patient, Ac-Powered devices

All Lift, Patient, Ac-Powered devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page