SW Bari Lift & Transfer, Model Number 38060000 recallZ-1595-2024 · Raye's Inc. · initiated Jan 25, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Raye's Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.

FDA root cause category: Device Design

What the company is doing

Sizewise Manufacturing issued an URGENT Voluntary Medical Device Recall to its consignees on 01/29/2024 by letter. The notice explained the issue, potential risk, and requested the discontinuation of use and disposal of the device. The firm sent a follow-up notice on 03/07/2024 asking the distributors to further notify their customers (users of the device).

Recall details

Recall number
Z-1595-2024
Status
Open, Classified
Initiated by the company
Jan 25, 2024
Posted by the FDA
Apr 18, 2024
Recalling firm
Raye's Inc.
Quantity
1,149 units
Product code
FNG
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, UK.

Product description

SW Bari Lift & Transfer, Model Number 38060000

Code information (models, lots, serial numbers)

UDI/DI 00845699001040, All serial numbers

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates