Skip to content

Latitude™ EV 0030313Recalls & reported problems

Tornier, Inc. · Humeral Stem - Coated · Class 2 · Implant

All models of Latitude EV (24)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00846832002764), received between Aug 30, 2022 and Mar 30, 2026 — classified as 2 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Limb Fracture1
  • Unspecified Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 30, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, A REMOVAL OF LATITUDE EV TOTAL ELBOW REPLACEMENT WAS PERFORMED FOR A PATIENT WHO HAD AN INFECTION. THE PATIENT HAD A REVISION SURGERY ON (B)(6) 2025. THE SURGEON ACCESSED THE JOINT AND REMOVED A SIGNIFICANT AMOUNT OF BLACK MATTER AND THEN PROCEEDED TO TAKE THE HUMERAL AND ULNAR STEM OUT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Unspecified Infection

    MAUDE report 24732933 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD AN OSTEOSUTURE ON THE HUMERAL PROSTHESIS DUE TO THEM FALLING ON THEIR LEFT ELBOW. THERE WAS A PERIPROSTHETIC COMPLEX FRACTURE OF THE OLECRANON AND A DISPLACEMENT OF THE HUMERAL IMPLANT. THE PATIENT UNDERWENT A REVISION SURGERY TO REMOVE THE STANDARD STEM AND REPLACE WITH A LONG STEM.

    Patient problems: Limb Fracture

    MAUDE report 15323520 ↗

Other PROSTHESIS, ELBOW, SEMI-CONSTRAINED, CEMENTED devices

All PROSTHESIS, ELBOW, SEMI-CONSTRAINED, CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page