TEMA Elbow 1590.15.010Recalls & reported problems
Limacorporate SPA · Axle Small · Class 2 · Implant
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08033390128865), received between May 24, 2023 and Dec 12, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Unstable” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Oct 2, 2024Open, Classified
Due to a potential difference in mechanical behavior between the implanted device and tool(s) used during revision surgery.
Z-0744-2025 · Root cause: Process control · FDA record ↗
- Jan 25, 2023Open, Classified
Due to the incorrect component, that is not cleared to be marketed in the U.S., being included in its elbow prosthesis.
Z-1261-2023 · Root cause: Employee error · FDA record ↗
Most reported problems
Device problems
- Unstable2
- No Apparent Adverse Event1
- Use of Device Problem1
Patient problems
- Joint Laxity2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 12, 2025Unstable; Unstable
CONVERSION SURGERY PERFORMED ON (B)(6) 2025 DUE TO ELBOW INSTABILITY. THE PRE-EXISTING IMPLANT WAS NOT REMOVED AS THERE WERE NO SIGNS OF BREAKAGE, BUT SURGEON IMPLANTED THE AXLE #SMALL (PART NUMBER: 1590.15.010, LOT: 2504642, STERILIZATION: 2500077) BETWEEN THE HUMERAL AND ULNAR COMPONENTS TO LINK THEM TO IMPROVE STABILITY. THE SURGERY WAS COMPLETED AS INTENDED. PREVIOUS SURGERY DATE AND LIST OF IMPLANTED COMPONENTS IS UNKNOWN. PATIENT IS A FEMALE, DATE OF BIRTH ON (B)(6) 1950. THE EVENT OCCURRED IN THE UNITED STATES.
Patient problems: Joint Laxity
MAUDE report 23789533 ↗InjuryMay 9, 2025No Apparent Adverse Event; No Apparent Adverse Event
A CONVERSION SURGERY TOOK PLACE ON (B)(6) 2025, TO LINK AN AXLE TO A CUSTOM PROMADE COMPONENT IN A TOTAL ELBOW ARTHROPLASTY. THE SURGEON DID NOT REMOVE THE SUSPECTED COMPONENT, AS THERE WERE NO SIGNS OF BREAKAGE OR FRAGMENTATION. THE SURGERY WAS COMPLETED AS INTENDED. THE DEVICE WAS INSPECTED PRIOR TO USE AND FOUND TO BE ACCEPTABLE. THE PREVIOUS SURGERY WAS PERFORMED ON (B)(6) 2025. PRODUCT IMPLANTED: AXLE #SMALL (PRODUCT CODE: 9159015010U03, LOT N.: 2435932, STERILIZATION N.: (B)(4)). PATIENT FEMALE, 78 YEARS OLD. EVENT OCCURRED IN THE UNITED STATES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21993936 ↗InjuryMay 24, 2023Unstable; Use of Device Problem; Unstable; Use of Device Problem
ELBOW REVISION SURGERY WAS PERFORMED ON (B)(6) 2020, TO IMPLANT AN AXLE - SMALL DEVICE, COMMERCIAL CODE: 1590.15.010 - LOT #: 1813818 - STER. #: (B)(6) TO STABILIZE LOOSE IMPLANTS. INITIAL SURGERY INVOLVED IMPLANTATION OF THE LISTED COMPONENTS BELOW. THE IFU STATES THAT THE COMPONENTS USED IN THE TEMA ELBOW SYSTEM ARE INTENDED FOR CEMENTED APPLICATIONS. THE SURGEON ELECTED TO USE AN OFF-LABEL USE USING AN UNCEMENTED APPLICATION. IMPLANTED COMPONENTS FROM INITIAL SURGERY: HUMERAL BODY - SMALL, COMMERCIAL CODE: 1550.15.011 - LOT #: 1814787 - STER. #: (B)(6); ULNAR LINER - SMALL, COMMERCIAL CODE: 1560.50.011 - LOT #: 18AT1RU - STER. #: (B)(6); HUMERAL STEM #H8, COMMERCIAL CODE: 1504.14.081 - LOT #: 1815323 - STER. #: (B)(6); ULNAR BODY-SMALL, COMMERCIAL CODE: 1552.14.011 - LOT #: 1814146 - STER. #: (B)(6); ULNAR STEM #U7, COMMERCIAL CODE: 1575.14.050 - LOT #: 1811263 - STER. #: (B)(6). INITIAL SURGERY: (B)(6) 2020. PATIENT: MALE. BIRTH DATE: (B)(6) 1945. APPROX. WEIGHT (INITIAL SURGERY): 79KG. ACTIVITY LEVEL: SEDENTARY. ADDITIONAL: POOR BONE QUALITY AND UNLINKED DURING INITIAL SURGERY. EVENT HAPPENED IN US.
Patient problems: Joint Laxity
MAUDE report 16990239 ↗
Other Prosthesis, Elbow, Semi-Constrained, Cemented devices
- TEMA Elbow 1590.15.020Limacorporate SPA · Prosthesis, Elbow, Semi-Constrained, Cemented1 recall1 report
- Latitude™ - Elbow DKY068Tornier, Inc. · PROSTHESIS, ELBOW, SEMI-CONSTRAINED, CEMENTED20 reports
- DURALOC 1241-46-526Depuy Orthopaedics, Inc. · PROSTHESIS, ELBOW, SEMI-CONSTRAINED, CEMENTED17 reports
- Latitude™ - Elbow DKY067Tornier, Inc. · PROSTHESIS, ELBOW, SEMI-CONSTRAINED, CEMENTED8 reports
- RHS - RADIAL PROSTHESIS DTJ302Tornier · PROSTHESIS, ELBOW, SEMI-CONSTRAINED, CEMENTED6 reports
- Latitude™ EV 0030221Tornier, Inc. · PROSTHESIS, ELBOW, SEMI-CONSTRAINED, CEMENTED5 reports
- Latitude™ - Elbow DKY082Tornier, Inc. · PROSTHESIS, ELBOW, SEMI-CONSTRAINED, CEMENTED5 reports
- CRF II - RADIAL PROSTHESIS DTJ022Tornier · PROSTHESIS, ELBOW, SEMI-CONSTRAINED, CEMENTED5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.