Skip to content

Tornier Latitude 9030200Recalls & reported problems

Tornier, Inc. · ULNAR JIG · Class 1

All models of Tornier Latitude (7)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00846832015160), received between Feb 16, 2022 and May 27, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam2

Patient problems

  • Insufficient Information2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 27, 2025Mechanical Jam; Mechanical Jam

    DURING PRIMARY RIGHT TOTAL ELBOW REPLACEMENT, SEVERAL ISSUES AROSE REGARDING LATITUDE EV INSTRUMENT MALFUNCTION. 1. DURING INITIAL PREPARATION OF THE ULNAR, THE ULNAR JIG, RIGHT WAS FITTED IN PLACE AS PER THE SURGICAL TECHNIQUE. THE LARGE SIZING HUMERAL SPOOL WAS SLID OVER THE CENTRAL AXEL OF THE JIG, AND THE LARGE BELLSAW WAS THEN PLACED OVER THE SPOOL, IN PREPARATION FOR PROFILING THE PROXIMAL ULNAR. WHEN THE SURGEON ACTIVATED THE DRILL-REAMER POWER TOOL, THE BELL SAW CENTRAL AXEL JAMMED WITHIN THE HOLLOW CENTRAL AXEL OF THE JIG. THE SURGEON WAS UNABLE TO ADVANCE THE BELLSAW, OR REMOVE IT, AS IT APPEARED THAT THE METAL SURFACES HAD JAMMED TOGETHER. THE JIG WAS FULLY LOOSENED, THE BELLSAW DETACHED FROM THE DRILL / REAMER POWER TOOL, AND THE JAMMED CONSTRUCT WAS REMOVED, CAREFULLY, FROM THE ELBOW JOINT. THE BELLSAW WAS THEN ABLE TO BE DETACHED FROM THE JIG. ON INSPECTION, THE INNER HOLLOW TUBE OF THE ULNAR JIG, RIGHT, APPEARED TO HAVE METAL FATIGUE, CAUSING THE CENTRAL AXEL OF THE BELLSAW TO JAM. THE SURGEON DECIDED TO USE THE ULNAR JIG, LEFT, FLIPPED IT OVER, POSITIONED IT IN PLACE, AND COMPLETED THE PROFILING OF THE PROXIMAL ULNAR, TO HIS SATISFACTION. NO FURTHER ISSUES WITH USIN

    Patient problems: Insufficient Information

    MAUDE report 22086850 ↗
  • MalfunctionFeb 16, 2022Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE ULNAR REAMER GOT STUCK AND WOULDN'T WORK.

    Patient problems: Insufficient Information

    MAUDE report 13546462 ↗

Other Guide, surgical, instrument devices

All Guide, surgical, instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page