Skip to content

Tornier Perform MWJ111Recalls & reported problems

Tornier, Inc. · Central Screw Drill · Class 1

Adverse event reports
10
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00846832062348), received between May 10, 2021 and May 8, 2026 — classified as 1 injury, 9 malfunctions.

The most frequently reported specific device problem is “Fracture” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture9
  • Break8

Patient problems

  • Foreign Body In Patient5
  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2026Fracture

    DRILL BIT BROKE DURING CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25128212 ↗
  • MalfunctionApr 28, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT A 6.5 CENTRAL DRILL BROKE DURING A PRIMARY SURGERY. NO OTHER INSTRUMENTS OR IMPLANTS WERE USED IN COMBINATION WITH THE AFFECTED INSTRUMENT. ALL DEBRIS WAS REMOVED FROM THE FIELD. THIS DRILL IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21922029 ↗
  • MalfunctionFeb 19, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED DURING INBOUND INSPECTION FOLLOWING A PRIMARY SURGERY THAT A 6.5MM CENTRAL SCREW DRILL TIP WAS BROKEN. THERE WAS NO REPORTED PRESENCE OF HARD BONE OR DEBRIS. THE SCREW DRILL HAD BEEN IN SERVICE FOR OVER FIVE YEARS. THE DEVICE WAS INSPECTED PRIOR TO BEING SENT TO THE CUSTOMER. A TRACKING NUMBER WAS PROVIDED FOR ITS RETURN TO THE MANUFACTURER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21412889 ↗
  • MalfunctionAug 22, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT A HANDLE BROKE OFF, AND THE TIP BROKE WHILE THE DOCTOR WAS DRILLING. ALL BROKEN PIECES WERE RECOVERED, AND THE ITEM WAS DISCARDED IMMEDIATELY AFTER THE INCIDENT.

    Patient problems: Insufficient Information

    MAUDE report 20029577 ↗
  • InjuryNov 1, 2023Break; Fracture; Break; Fracture

    AS REPORTED: "THE PATIENT WAS UNDERGOING A PERFORM STEM AND PERFORM REVERSED TOTAL SHOULDER PROSTHESIS, HAD A HIGH BONE DENSITY AND A SCLEROTIC GLENOID, WHICH HAD PREVIOUSLY TWISTED THE THREADED SPINDLE, FOR GLENOID AIMING GUIDANCE. DURING DRILLING OF THE CENTRAL SCREW OF THE PERFORM REVERSED BASEPLATE, THE 6.5-DIAMETER WIRE, REF MWJ111, BROKE OFF, LEAVING 2/3 OF IT IN THE PATIENT'S GLENOID. THE SURGEON WAS UNABLE TO REMOVE THE DEBRIS, AND INSERTED A SHORT 7MM STUD IN PLACE OF THE 6.5X25 CENTRAL SCREW INITIALLY CHOSEN. THE OPERATION WAS COMPLETED WITHOUT DIFFICULTY."

    Patient problems: Foreign Body In Patient

    MAUDE report 18047188 ↗
  • MalfunctionSep 19, 2023Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE DRILL BROKE OFF INSIDE THE PATIENT AND COULD NOT RETRIEVED.

    Patient problems: Foreign Body In Patient

    MAUDE report 17777293 ↗
  • MalfunctionJun 29, 2022Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE CENTRAL SCREW BROKEN DURING SURGERY. THE CUSTOMER HAD A DRILL FROM ANOTHER COMPANY AVAILABLE AS A SUBSTITUTE TO SUCCESSFULLY COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14851278 ↗
  • MalfunctionMar 3, 2022Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT DURING THE PROCEDURE THE SURGEON DRILLED FOR 6.5MM CENTRAL SCREW AND NOTICED THE END OF THE DRILL WAS BROKEN AFTER USE. THE TIP OF THE DRILL WAS LODGED IN THE BONE BUT THE SCREW PLACEMENT WAS SUCCESSFUL.

    Patient problems: Foreign Body In Patient

    MAUDE report 13659707 ↗
  • MalfunctionOct 6, 2021Fracture; Fracture

    IT WAS REPORTED THAT THE DISTAL PART OF THE DRILL FOR THE CENTRAL SCREW BROKE OFF DURING GLENOID PREPARATION LEAVING A PART WITHIN THE PATIENT'S GLENOID.

    Patient problems: Foreign Body In Patient

    MAUDE report 12585236 ↗
  • MalfunctionMay 10, 2021Break; Break

    ALLEGEDLY, THE SURGEON IDENTIFIED THROUGH X-RAY THAT THERE WAS SOMETHING IN THE PATIENT'S BONE RESEMBLING A TIP OF A DRILL BIT. UPON A SEARCH IN STERILE PROCESSING THE DRILL WAS LOCATED AND THE DRILL IS MISSING THE TOP PORTION.

    Patient problems: Foreign Body In Patient

    MAUDE report 11803706 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.