Tornier Perform MWJ127Recalls & reported problems

Tornier, Inc. · Peripheral and Sphere Screwdriver Bit · Class 1

All models of Tornier Perform (40)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Tornier Perform screwdriver bit has 129 adverse event reports on record, with no deaths and 3 injuries. The reports mostly describe the bit fracturing or its tip breaking off while tightening screws, and a few mention fragments left in the patient. Most say no clinical signs were reported. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
114
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 114 adverse event reports that list this exact device (DI 00846832062492), received between May 6, 2021 and Sep 29, 2026 — classified as 3 injuries, 111 malfunctions.

The most frequently reported specific device problem is “Fracture” (111 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 6 reportsDec 2024: 11 reportsJan 2025: 2 reportsFeb 2025: 8 reportsMar 2025: 5 reportsApr 2025: 7 reportsMay 2025: 4 reportsJun 2025: 3 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 6 reportsOct 2025: 5 reportsNov 2025: 6 reportsDec 2025: 5 reportsJan 2026: 1 reportsFeb 2026: 4 reportsMar 2026: 5 reportsApr 2026: 2 reportsMay 2026: 10 reportsJun 2026: 3 reportsJul 2026: 4 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture111
  • Break65
  • Material Twisted/Bent7
  • Material Deformation2
  • Material Erosion1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions108
  • Insufficient Information6
  • Foreign Body In Patient3

Recent adverse event reports

The 25 most recent of 114 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Fracture

    Patient problems: Insufficient Information

    MAUDE report 50935307 ↗
  • MalfunctionSep 17, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50809706 ↗
  • MalfunctionJul 29, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50074249 ↗
  • MalfunctionJul 27, 2026Fracture

    As reported: "Found tip of driver to be sheared off."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50046410 ↗
  • MalfunctionJul 9, 2026Fracture; Material Deformation

    Screwdriver bits were warped or broken.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25764165 ↗
  • MalfunctionJul 7, 2026Fracture; Material Twisted/Bent

    IN A REVISION, THE SURGEON WAS TRYING THE ADVANCE THE CENTRAL SCREW IT THE POLY USING BOTH DRIVERS. IN THE PROCESS THE BOTH DRIVERS WERE BROKE/BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25737265 ↗
  • InjuryJun 9, 2026Fracture

    DURING A REVERSE SHOULDER ARTHROPLASTY THE SCREWDRIVER TIP (MWJ127) BROKE AND SNAPPED OFF INSIDE THE GLENOSPHERE WHEN TIGHTENING THE GLENOSPHERE SCREW. THE TIP WAS BURIED AND STUCK IN THE SCREW HEAD OF THE GLENOSPHERE SO REMAINED IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 25446813 ↗
  • MalfunctionJun 9, 2026Fracture

    T20 DRIVER TIP BROKE OFF IN PERFORM SPHERE EXTRACTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25446809 ↗
  • MalfunctionJun 2, 2026Fracture

    IT WAS REPORTED THAT THE DRIVER BROKE DURING A CASE AND IS UNUSABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25362349 ↗
  • MalfunctionMay 13, 2026Fracture

    IT WAS REPORTED THAT DURING A PRIMARY SURGERY THE TORNIER DRIVER FOR THE RATCHETING HANDLES HIP BROKE OFF WHILE INSERTING A SCREW, EVERYTHING WAS RECOVERED NO ISSUES. THE REMAINING SCREWS WERE INSERTED WITH ANOTHER SCREWDRIVER. THE DURATION OF THE SURGICAL DELAY IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25162473 ↗
  • MalfunctionMay 12, 2026Fracture

    DRIVER TIP BROKE DURING PROCEDURE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25151401 ↗
  • MalfunctionMay 12, 2026Fracture

    DRIVER TIP BROKE DURING PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25151400 ↗
  • MalfunctionMay 12, 2026Fracture

    DRIVER TIP BROKE DURING PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25151390 ↗
  • MalfunctionMay 12, 2026Fracture

    DRIVER TIP BROKE DURING PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150722 ↗
  • MalfunctionMay 12, 2026Fracture

    DRIVER TIP BROKE DURING PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150678 ↗
  • MalfunctionMay 12, 2026Fracture

    DRIVER TIP BROKE DURING PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150677 ↗
  • MalfunctionMay 12, 2026Fracture

    DRIVER TIP BROKE DURING PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25150527 ↗
  • MalfunctionMay 5, 2026Fracture

    IT WAS REPORTED THAT THE SCREWDRIVER TIP WAS BROKEN DURING PLACEMENT OF PERIPHERAL SCREWS WITH RATCHETING HANDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25088740 ↗
  • MalfunctionMay 1, 2026Fracture

    IT WAS REPORTED THAT THE DRIVER TIP BROKE WHILE TIGHTENING SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25056844 ↗
  • MalfunctionApr 28, 2026

    DURING THE PROCEDURE A PIECE OF THE SCREW BIT BROKE OFF IN THE PATIENT'S LEFT SHOULDER. MD PULLED OUT THE BROKEN PIECE. X RAY CAME TO THE ROOM WITH A C ARM TO GET X RAY. MD CONFIRMED THAT NO PIECES WERE LEFT IN THE PATIENT'S ARM UTILIZING X RAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25014472 ↗
  • MalfunctionApr 23, 2026Fracture

    SALES REPRESENTATIVE REPORTED THAT THE SCREWDRIVER BIT BROKE DURING THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24968381 ↗
  • MalfunctionMar 31, 2026Fracture

    SCREWDRIVER TIP BROKE WHEN THE SURGEON WAS INSERTING THE GLENOSPHERE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24745396 ↗
  • MalfunctionMar 20, 2026Fracture

    SCREWDRIVER SHAFT BROKE DURING GLENOSPHERE TIGHTENING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24656052 ↗
  • MalfunctionMar 19, 2026Fracture

    2 SCREWDRIVER TIPS WERE BROKEN INSERTING PERIPHERAL SCREWS ONTO THE BASEPLATE, AND UPON WASHING THE ATTACHMENT TO THE IM CUT GUIDE, IT WAS NOTICED THAT THE KNOB WAS BENT AND COULD NO LONGER BE SCREWED DOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24639271 ↗
  • MalfunctionMar 19, 2026Fracture

    2 SCREW DRIVER TIPS WERE BROKEN INSERTING PERIPHERAL SCREWS ONTO THE BASEPLATE, AND UPON WASHING THE ATTACHMENT TO THE IM CUT GUIDE, IT WAS NOTICED THAT THE KNOB WAS BENT AND COULD NO LONGER BE SCREWED DOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24638786 ↗