NA 4200355042Recalls & reported problems

Stryker Corporation · 4.2mm Radiolucent Drill Bit · Class 1

Quick summary

This Stryker 4.2mm radiolucent drill bit is used in surgical drilling. Most of the 36 reports describe the bit fracturing or breaking during procedures, sometimes leaving a fragment in the patient; 2 reports describe injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 04546540348593), received between Mar 26, 2018 and Apr 22, 2026 — classified as 1 injury, 33 malfunctions.

The most frequently reported specific device problem is “Fracture” (29 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture29
  • Break4
  • Delivered as Unsterile Product2

Patient problems

  • No Clinical Signs, Symptoms or Conditions20
  • No Consequences Or Impact To Patient10
  • Device Embedded In Tissue or Plaque8
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2026Fracture

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1 ¿ MARCH 31 2026. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 24954290 ↗
  • MalfunctionJan 23, 2026Fracture

    NO CHANGE.

    Patient problems: Device Embedded In Tissue or Plaque; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24163432 ↗
  • MalfunctionOct 28, 2025Fracture; Fracture

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399810 ↗
  • MalfunctionJul 29, 2025Fracture; Fracture

    THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 22647323 ↗
  • MalfunctionMay 23, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A PROCEDURE THE TIP OF THE DRILL BIT BROKE. IT WAS ALSO REPORTED THAT THE BROKEN PIECE REMAINED IN THE PATIENT. IT WAS FURTHER REPORTED, THAT THERE WAS NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT, AND THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 22078579 ↗
  • MalfunctionApr 21, 2025Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876047 ↗
  • MalfunctionJan 21, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS REPORTEDLY RECEIVED IN SUCH A MANNER TO INDICATE THAT ITS STERILITY HAD BEEN COMPROMISED. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21200734 ↗
  • MalfunctionOct 23, 2023Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17985865 ↗
  • MalfunctionAug 7, 2023Delivered as Unsterile Product; Delivered as Unsterile Product

    THE USER FACILITY REPORTED THAT THE PACKAGING WAS OBSERVED TO BE UNSEALED PRIOR TO OPENING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17468851 ↗
  • MalfunctionJul 26, 2023Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 17398737 ↗
  • MalfunctionJan 24, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING AN OPEN REDUCTION WITH INTERNAL FIXATION OF NON-UNION INTERTROCHANTERIC FRACTURE SURGERY; THE TIP OF THE DRILL BIT BROKE. IT WAS ALSO REPORTED THAT THE BROKEN PIECE REMAINED IN THE PATIENT. IT WAS FURTHER REPORTED THAT THERE WAS NO MEDICAL INTERVENTION, WITH NO CLINICALLY-SIGNIFICANT SURGICAL DELAY AS A RESULT OF THIS EVENT, AND THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 16231753 ↗
  • MalfunctionJan 12, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A INTRAMEDULLARY NAIL SURGERY FOR FEMORAL DIAPHYSEAL FRACTURE; WHILE DRILLING TO INSERT THE SCREW INTO THE AP HOLE, THE DRILL BIT BROKE. IT WAS ALSO REPORTED THAT THE BROKEN PIECE REMAINED IN THE PATIENT. IT WAS ALSO REPORTED THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO FURTHER INFORMATION REPORTED.

    Patient problems: Device Embedded In Tissue or Plaque; Insufficient Information

    MAUDE report 16150021 ↗
  • MalfunctionDec 8, 2022Fracture; Fracture

    THE USER FACILITY REPORTED THAT THE WELD WAS FRACTURED ON THE HOUSING DURING A PROCEDURE. THE BROKEN HOUSING COULD CAUSE THE NON-STERILE BATTERY TO BE EXPOSED TO THE STERILE FIELD. THERE WERE NO ADVERSE CONSEQUENCES. ATTEMPTS ARE BEING MADE TO OBTAIN ADDITIONAL INFORMATION ABOUT THE EVENT; NO FURTHER INFORMATION WAS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15938747 ↗
  • MalfunctionDec 8, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A PROCEDURE, THE DRILL BROKE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15938743 ↗
  • MalfunctionAug 2, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE RADIOLUCENT DRILL BIT BROKE, THE BROKEN PIECES WERE REMOVED FROM THE PATIENT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15152837 ↗
  • MalfunctionJul 14, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING AN AP HOLE OF THE T2ALPHA TIBIA SURGERY, THE DRILL BIT BROKE. IT WAS ALSO REPORTED THAT THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT¿S BONE. IT WAS FURTHER REPORTED THAT THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15021594 ↗
  • InjuryJul 6, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THE DRILL BIT BROKE. IT WAS ALSO REPORTED THAT THE FRACTURED TIP OF THE DRILL BIT WAS STILL STUCK IN THE BONE. IT WAS FURTHER REPORTED THAT THE BONE WAS SHAVED AROUND THE DRILL TIP SO THE FRAGMENT COULD BE REMOVED SUCCESSFULLY. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14938705 ↗
  • MalfunctionJun 20, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING T2 SCNL SURGERY, IT WAS NOTED THAT THE DRILL BIT TIP BROKE WHILE DRILLING INTO THE PROXIMAL TRUE CIRCLE HOLE AND WAS REMOVED AS PART OF THE PROCEDURE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14748256 ↗
  • MalfunctionJun 3, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A T2SCN SURGERY; WHILE DRILLING THE PROXIMAL SCREW HOLE OF NAIL, THE DRILL BIT BROKE. IT WAS ALSO REPORTED THAT THE BROKEN PIECE REMAINED IN THE PATIENT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Device Embedded In Tissue or Plaque; No Clinical Signs, Symptoms or Conditions

    MAUDE report 14593989 ↗
  • MalfunctionMar 25, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A T2SCN LONG NAIL SURGERY; WHILE DRILLING THE PROXIMAL SCREW HOLE OF NAIL, THE DRILL BIT BROKE. IT WAS ALSO REPORTED THAT THE BROKEN PIECE REMAINED IN THE PATIENT. IT WAS FURTHER REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13902074 ↗
  • MalfunctionJan 4, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THE DRILL BIT BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13155874 ↗
  • MalfunctionSep 29, 2021Fracture; Fracture

    IT WAS REPORTED THAT DURING A T2 SCN SURGERY; WHILE DRILLING THE PROXIMAL SCREW HOLE OF NAIL, THE DRILL BIT BROKE. NO FURTHER INFORMATION WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12548556 ↗
  • MalfunctionAug 27, 2021Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE DRILL BIT BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12380270 ↗
  • MalfunctionDec 17, 2020Fracture; Fracture

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE ATTACHMENT STOPPED AND THE DRILL BIT BROKE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11032141 ↗
  • MalfunctionAug 31, 2020Fracture; Fracture

    IT WAS REPORTED THAT DURING A FRACTURE REPAIR OF DISTAL FEMUR WITH NAIL, TWO RADIOLUCENT DRILL BITS BROKE AT THE CUTTING END. IT WAS ALSO REPORTED THAT THE DRILL FRAGMENT IS COMPLETELY BURIED IN THE BONE AND CANNOT BE REMOVED. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.THIS REPORT IS FOR THE SECOND DRILL BIT THAT BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10471997 ↗

Other BIT, SURGICAL devices

All BIT, SURGICAL devices →