AEQUALIS™ FLEX REVIVE™ ARS742501Recalls & reported problems
Tornier, Inc. · Class 2
All models of AEQUALIS FLEX REVIVE (24)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00846832068814), received between Dec 1, 2023 and Apr 16, 2026 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture4
- Break3
- Material Deformation2
- Detachment of Device or Device Component1
- Device Difficult to Program or Calibrate1
- Device Difficult to Setup or Prepare1
- Material Protrusion/Extrusion1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 16, 2026Fracture; Material Protrusion/Extrusion
BOTTOM HALF OF TRIAL HAS EXPANDED AND CANNOT BE USED ANYMORE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24896489 ↗MalfunctionJul 25, 2025Device Difficult to Program or Calibrate; Material Deformation; Device Difficult to Program or Calibrate; Material Deformation
IT WAS REPORTED THAT A 9MM PROXIMAL BODY TRIAL NO LONGER TIGHTLY SECURES TO THE TRIAL STEMS OR TRIAL SPACERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22613492 ↗MalfunctionApr 21, 2025Break; Fracture; Break; Fracture
IT WAS REPORTED THAT WHILE IMPACTING THE TRIAL ASSEMBLY INTO THE HUMERUS, THE PROXIMAL BODY DETACHED FROM THE STEM DUE TO BROKEN RETAINING TABS. THE STEM TRIAL REMAINED IN THE HUMERAL CANAL AND WAS REMOVED USING NEEDLE NOSE PLIERS. ALL DEBRIS WAS CONFIRMED REMOVED VIA X-RAY. THE CASE WAS DELAYED APPROXIMATELY 40 MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21877653 ↗MalfunctionDec 27, 2024Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare
AS REPORTED: "ONE OF THE 9X20 SPACERS WOULD NOT LOCK ONTO STEM. WE USED THE SECOND ONE IN THE SET SO DID NOT INTERRUPT THE CASE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21030803 ↗MalfunctionOct 14, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE PROXIMAL BODY TRIAL WAS BENT AT THE BOTTOM DURING A CASE. A BACKUP DEVICE WAS USED TO COMPLETE THE SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20441038 ↗MalfunctionOct 8, 2024Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING A TRAINING LAB SESSION AT THE ST. LOUIS HUB, THE TAB ON THE BOTTOM OF A SIZE 9 REVIVE PROXIMAL BODY BENT DURING ASSEMBLY. WHEN AN ATTEMPT WAS MADE TO STRAIGHTEN THE TAB, IT BROKE OFF COMPLETELY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20397362 ↗InjuryDec 1, 2023Break; Fracture; Detachment of Device or Device Component; Material Deformation; Naturally Worn; Break; Fracture; Detachment of Device or Device Component; Material Deformation; Naturally Worn
DURING A RECENT SURGERY FOR A TOTAL REVERSE FRACTURE CASE, A PORTION OF THE 9X20 TRIAL SPACER FROM THE TORNIER FLEX REVIVE HUMERAL INSTRUMENTS TRAY FRACTURED AND BECAME STUCK INSIDE THE HUMERAL CANAL. THIS INCIDENT CAUSED A SIGNIFICANT DELAY IN THE SURGERY. THE SURGICAL TEAM HAD TO REMOVE A SECTION OF THE HUMERUS TO ACCESS AND REMOVE THE TRAPPED TRIAL STEM, WHICH WAS LATER REATTACHED USING CERCLAGE. THE INCIDENT RESULTED IN A 3-HOUR SURGICAL DELAY, BUT ALL DEBRIS WAS SUCCESSFULLY REMOVED USING A BUR INSTRUMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18251081 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.