Skip to content

AEQUALIS™ FLEX REVIVE™ ARS742501Recalls & reported problems

Tornier, Inc. · Class 2

All models of AEQUALIS FLEX REVIVE (24)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00846832068814), received between Dec 1, 2023 and Apr 16, 2026 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Fracture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture4
  • Break3
  • Material Deformation2
  • Detachment of Device or Device Component1
  • Device Difficult to Program or Calibrate1
  • Device Difficult to Setup or Prepare1
  • Material Protrusion/Extrusion1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 16, 2026Fracture; Material Protrusion/Extrusion

    BOTTOM HALF OF TRIAL HAS EXPANDED AND CANNOT BE USED ANYMORE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24896489 ↗
  • MalfunctionJul 25, 2025Device Difficult to Program or Calibrate; Material Deformation; Device Difficult to Program or Calibrate; Material Deformation

    IT WAS REPORTED THAT A 9MM PROXIMAL BODY TRIAL NO LONGER TIGHTLY SECURES TO THE TRIAL STEMS OR TRIAL SPACERS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22613492 ↗
  • MalfunctionApr 21, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT WHILE IMPACTING THE TRIAL ASSEMBLY INTO THE HUMERUS, THE PROXIMAL BODY DETACHED FROM THE STEM DUE TO BROKEN RETAINING TABS. THE STEM TRIAL REMAINED IN THE HUMERAL CANAL AND WAS REMOVED USING NEEDLE NOSE PLIERS. ALL DEBRIS WAS CONFIRMED REMOVED VIA X-RAY. THE CASE WAS DELAYED APPROXIMATELY 40 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21877653 ↗
  • MalfunctionDec 27, 2024Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    AS REPORTED: "ONE OF THE 9X20 SPACERS WOULD NOT LOCK ONTO STEM. WE USED THE SECOND ONE IN THE SET SO DID NOT INTERRUPT THE CASE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21030803 ↗
  • MalfunctionOct 14, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE PROXIMAL BODY TRIAL WAS BENT AT THE BOTTOM DURING A CASE. A BACKUP DEVICE WAS USED TO COMPLETE THE SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20441038 ↗
  • MalfunctionOct 8, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT DURING A TRAINING LAB SESSION AT THE ST. LOUIS HUB, THE TAB ON THE BOTTOM OF A SIZE 9 REVIVE PROXIMAL BODY BENT DURING ASSEMBLY. WHEN AN ATTEMPT WAS MADE TO STRAIGHTEN THE TAB, IT BROKE OFF COMPLETELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20397362 ↗
  • InjuryDec 1, 2023Break; Fracture; Detachment of Device or Device Component; Material Deformation; Naturally Worn; Break; Fracture; Detachment of Device or Device Component; Material Deformation; Naturally Worn

    DURING A RECENT SURGERY FOR A TOTAL REVERSE FRACTURE CASE, A PORTION OF THE 9X20 TRIAL SPACER FROM THE TORNIER FLEX REVIVE HUMERAL INSTRUMENTS TRAY FRACTURED AND BECAME STUCK INSIDE THE HUMERAL CANAL. THIS INCIDENT CAUSED A SIGNIFICANT DELAY IN THE SURGERY. THE SURGICAL TEAM HAD TO REMOVE A SECTION OF THE HUMERUS TO ACCESS AND REMOVE THE TRAPPED TRIAL STEM, WHICH WAS LATER REATTACHED USING CERCLAGE. THE INCIDENT RESULTED IN A 3-HOUR SURGICAL DELAY, BUT ALL DEBRIS WAS SUCCESSFULLY REMOVED USING A BUR INSTRUMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18251081 ↗

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page