Skip to content

AEQUALIS™ FLEX REVIVE™ ARS342007Recalls & reported problems

Tornier, Inc. · Class 2 · Implant

All models of AEQUALIS FLEX REVIVE (24)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00846832069347), received between Jul 29, 2025 and Jul 29, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Biocompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Biocompatibility1
  • Loss of Osseointegration1

Patient problems

  • Inadequate Osseointegration1
  • Metal Related Pathology1
  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 29, 2025Loss of Osseointegration; Biocompatibility; Loss of Osseointegration; Biocompatibility

    A REVISION SURGERY WAS REQUIRED FOLLOWING THE DISENGAGEMENT OF THE TRAY FROM THE REVIVE STEM IN AN ACTIVE, INDEPENDENTLY MOBILE PATIENT WHO PRESENTED WITH ACUTE SHOULDER PAIN. INTRAOPERATIVE FINDINGS REVEALED A LOOSE STEM, FAILURE AT THE MORSE TAPER INTERFACE, PSEUDOMEMBRANE FORMATION AROUND THE TRAY, AND SIGNIFICANT METALLOSIS. THE STEM, TRAY, AND LINER WERE REVISED USING A LARGER STEM SIZE, WHILE THE GLENOSPHERE AND SCREWS REMAINED INTACT DUE TO STABLE FIXATION. POSTOPERATIVELY, THE PATIENT UNDERWENT STANDARD REHABILITATION, INCLUDING SLING IMMOBILIZATION FOR 3 TO 4 WEEKS FOLLOWED BY GENTLE MOBILIZATION UNDER PHYSIOTHERAPY SUPERVISION. X-RAY IMAGES ARE NOT INCLUDED DUE TO PATIENT CONFIDENTIALITY.

    Patient problems: Inadequate Osseointegration; Metal Related Pathology; Pain

    MAUDE report 22649365 ↗

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates