Tornier Perform MWM054Recalls & reported problems

Tornier, Inc. · Humeral Head Impactor Tip · Class 1

All models of Tornier Perform (38)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Tornier Perform humeral head impactor tip is used to seat the shoulder ball implant during surgery. The 52 reports all describe the metal tip fracturing, cracking or shattering during impaction, sometimes leaving fragments in the surgical field and in one case breaking after the seventh impact. No deaths or injuries were reported, and there are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
49
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 00846832082964), received between Jun 22, 2022 and Jul 15, 2026 — classified as 49 malfunctions.

The most frequently reported specific device problem is “Fracture” (42 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 3 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 5 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 4 reportsSep 2025: 2 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture42
  • Break36
  • Crack10
  • Material Fragmentation6
  • Naturally Worn2

Patient problems

  • No Clinical Signs, Symptoms or Conditions49
  • Insufficient Information2

Recent adverse event reports

The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 15, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25829139 ↗
  • MalfunctionJul 6, 2026Fracture

    PRODUCT IS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25721812 ↗
  • MalfunctionNov 24, 2025Fracture; Fracture

    IT WAS REPORTED THAT ON THE 7TH IMPACT IT JUST BROKE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23638942 ↗
  • MalfunctionOct 28, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE IMPACTOR TIP BROKE DURING IMPACTION OF THE GLENOSPHERE DURING A PRIMARY SURGERY. DEBRIS WAS REMOVED FROM THE FIELD, THE GLENOSPHERE DIDN'T NEED ANY MORE IMPACTION. SURGERY CONTINUED AFTER LOOKING AT THE GLENOSPHERE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23401609 ↗
  • MalfunctionOct 8, 2025Fracture; Fracture

    IMPACTOR TIP BROKE INTO TWO PIECES DURING SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23245306 ↗
  • MalfunctionSep 11, 2025Break; Fracture; Break; Fracture

    BROKEN IMPACTOR TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23028249 ↗
  • MalfunctionSep 10, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT BOTH INSTRUMENTS DID NOT PERFORM DUE TO BREAKING FROM USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23020490 ↗
  • MalfunctionAug 27, 2025Crack; Fracture; Crack; Fracture

    IT WAS REPORTED THAT THE IMPACTOR TIP WAS FOUND BROKEN IN HALF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22909483 ↗
  • MalfunctionAug 14, 2025Crack; Crack

    HEAD IMPACTION TIP WAS CRACKED DURING IMPACTION IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22805350 ↗
  • MalfunctionAug 14, 2025Break; Fracture; Break; Fracture

    THE GLENOSPHERE IMPACTOR TIP WAS DAMAGED DURING IMPACTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22805310 ↗
  • MalfunctionAug 12, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THERE WAS A BREAKAGE DURING THE CASE AND A BACK UP WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22784870 ↗
  • MalfunctionJul 10, 2025Crack; Crack

    IT WAS REPORTED THAT THE IMPACTOR TIP FROM PERFORM HUMERAL CORE TRAY CRACKED WHILE IMPACTING THE GLENOSPHERE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22469800 ↗
  • MalfunctionJun 18, 2025Break; Fracture; Break; Fracture

    THE HEAD IMPACTOR TIP BROKE WHILE IMPACTING DURING PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22288108 ↗
  • MalfunctionMay 8, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE IMPACTOR TIP BROKE INTRA-OP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21989022 ↗
  • MalfunctionApr 30, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE IMPACTOR TIP BROKE DURING THE INSTALLATION OF THE GLENOSPHERE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21936760 ↗
  • MalfunctionApr 28, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE WHITE HUMERAL HEAD IMPACTOR TIP BROKE DURING IMPACTION OF GLENOSPHERE DURING CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21922302 ↗
  • MalfunctionMar 28, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT IMPACTOR TIP WAS BROKEN DURING MED. PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21726603 ↗
  • MalfunctionMar 19, 2025Break; Fracture; Break; Fracture

    AS REPORTED: "IMPACTION TIP BROKE DURING CASE WHILE IMPACTING WITH A MALLET. ALL PIECES WERE EXTRACTED"

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21647441 ↗
  • MalfunctionMar 19, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE IMPACTOR TIP BROKE IN HALF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21642219 ↗
  • MalfunctionMar 17, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT DURING SURGERY BOTH IMPACTOR TIPS BROKE DURING IMPACTION. ALL DEBRIS WAS REMOVED FROM THE SURGICAL FIELD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21626165 ↗
  • MalfunctionMar 14, 2025Crack; Crack

    IT WAS REPORTED THAT THE PEG BROKE OFF OF THE TRIAL AND THE IMPACTOR TIP BROKE IN HALF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21610769 ↗
  • MalfunctionFeb 6, 2025Break; Fracture; Break; Fracture

    THE GLENOSPHERE WAS BROKEN AFTER IT WAS STRUCK. IT WAS REPORTED THAT: "THE DEFECT WAS ONLY NOTICED AFTER THE PROCEDURE.".

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21328207 ↗
  • MalfunctionJan 27, 2025Break; Crack; Fracture; Break; Crack; Fracture

    IT WAS REPORTED THAT DURING A PRIMARY SURGERY, THE IMPACTOR TIP BROKE OR CRACKED IN HALF WHILE IMPACTING THE SPHERE IMPLANT USING THE SPHERE IMPACTOR TIP. NO DEBRIS WAS REPORTED, AS THE DEVICE BROKE INTO TWO PIECES, WHICH WERE REMOVED FROM THE FIELD. THE SURGERY WAS SUCCESSFULLY COMPLETED USING A DIFFERENT IMPACTOR TIP. THE REPORTER STATED THAT THEY TAPPED THE SCREWDRIVER TO ENGAGE SCREW OF SPHERE AND THAT ALSO HELPED/WORKED. THE COMPLAINT DEVICE WAS NOT AVAILABLE FOR EVALUATION BY THE MANUFACTURER, AS IT WAS REPORTED TO BE LOST. IT WAS APPROXIMATED THAT THE DEVICE HAS BEEN IN SERVICE FOR 2-3 YEARS, AND HAS BEEN USED AND REPROCESSED OVER 100 TIMES. THE DEVICE WAS INSPECTED PRIOR TO BEING SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21247654 ↗
  • MalfunctionJan 3, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE SURGEON WAS IMPLANTING A PERFORM GLENOSPHERE USING THE PERFORM HUMERAL IMPACTOR TIP FROM THE PERFORM HUMERAL CORE TRAY. DURING THE PROCEDURE, THE PLASTIC TIP OF THE IMPACTOR SPLIT IN TWO AFTER ONE STRIKE WITH A MALLET. ONE HALF STAYED IN THE JOINT, AND THE OTHER REMAINED THREADED INTO THE IMPACTOR HANDLE. BOTH PIECES WERE PROMPTLY RETRIEVED, INSPECTED, AND ACCOUNTED FOR, ENSURING NO FRAGMENTS WERE LEFT IN THE SURGICAL SITE. THE SURGERY CONTINUED WITHOUT DELAY USING AN ALTERNATE IMPACTOR TIP FROM THE PERFORM PMI TRAY, AND THE SPHERE LOCKING SCREW WAS SECURED PER TECHNIQUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21067827 ↗
  • MalfunctionDec 26, 2024Crack; Crack

    IT WAS REPORTED THAT THE IMPACTOR TIP IS CRACKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21021405 ↗

FDA clearance and classification

The FDA regulates product code HWA (impactor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Tornier Perform MWM054

Has the Tornier Perform MWM054 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00846832082964), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Tornier Perform MWM054?

The FDA's MAUDE database lists 49 adverse event reports for this device, received between Jun 22, 2022 and Jul 15, 2026. In those reports, the most common specific device problems are “Fracture” (42), “Break” (36) and “Crack” (10). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Tornier Perform MWM054?

None of the reports are classified as deaths, 0 as injuries and 49 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Tornier Perform MWM054 FDA approved?

Devices of this type (product code HWA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Tornier, Inc.

Who makes the Tornier Perform MWM054, and is it still sold?

The FDA's device database lists Tornier, Inc. as the company responsible for it (brand Tornier Perform, model MWM054). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other IMPACTOR devices

All IMPACTOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page