TORNIER PERFORM™ HUMERAL SYSTEM DWP2396Recalls & reported problems
Tornier, Inc. · Class 2 · Implant
All models of TORNIER PERFORM HUMERAL SYSTEM (68)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00846832084500), received between Nov 10, 2021 and Nov 10, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Difficult to Insert” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Difficult to Insert1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 10, 2021Difficult to Insert; Difficult to Insert
THE SURGEON IMPLANTED THE PERFORM HUMERAL #3 STANDARD STEM IN THE PATIENT AND WHILE ATTEMPTING TO IMPACT THE 39MM + 6 REVERSE INSERT HE COULDN¿T SUCCESSFULLY SEAT IT INTO THE STEM. AFTER SEVERAL ATTEMPTS IT WAS DECIDED TO OPEN A SECOND REVERSE INSERT. PRIOR TO ATTEMPTING TO IMPLANT THE SECOND REVERSE INSERT THE HUMERAL CUT SURFACE WAS INSPECTED, TRIMMED & REAMED TO ASSURE THAT THERE WAS NOTHING BLOCKING THE INSERT TO SEAT CORRECTLY. ON THE SECOND REVERSE IMPLANT ATTEMPT THE SURGEON EXPERIENCED THE SAME ISSUES AND ENDED UP DROPPING THE SECOND IMPLANT TO THE FLOOR. THE SURGEON THEN EXPLANTED THE PERFORM HUMERAL STEM AND ATTEMPTED SEVERAL TIMES TO IMPACT THE REVERSE INSERT THAT WAS OPENED THE FIRST TIME ON THE BACK TABLE WITH THE IMPACTION STAND WITHOUT SUCCESS. AT THAT TIME, HE ASKED FOR A NEW PERFORM HUMERAL STEM TO BE OPENED AND REVERSE INSERT. THE NEW PERFORM HUMERAL STEM AND INSERT WAS IMPLANTED WITHOUT ANY PROBLEMS. THIS WHOLE TROUBLESHOOTING PROCESS PROBABLY ADDED 20-30 MINUTES TO THE SURGICAL TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12787414 ↗MalfunctionNov 10, 2021Positioning Failure; Positioning Failure
THE SURGEON IMPLANTED THE PERFORM HUMERAL #3 STANDARD STEM IN THE PATIENT AND WHILE ATTEMPTING TO IMPACT THE 39MM + 6 REVERSE INSERT HE COULDN¿T SUCCESSFULLY SEAT IT INTO THE STEM. AFTER SEVERAL ATTEMPTS IT WAS DECIDED TO OPEN A SECOND REVERSE INSERT. PRIOR TO ATTEMPTING TO IMPLANT THE SECOND REVERSE INSERT THE HUMERAL CUT SURFACE WAS INSPECTED, TRIMMED & REAMED TO ASSURE THAT THERE WAS NOTHING BLOCKING THE INSERT TO SEAT CORRECTLY. ON THE SECOND REVERSE IMPLANT ATTEMPT THE SURGEON EXPERIENCED THE SAME ISSUES AND ENDED UP DROPPING THE SECOND IMPLANT TO THE FLOOR. THE SURGEON THEN EXPLANTED THE PERFORM HUMERAL STEM AND ATTEMPTED SEVERAL TIMES TO IMPACT THE REVERSE INSERT THAT WAS OPENED THE FIRST TIME ON THE BACK TABLE WITH THE IMPACTION STAND WITHOUT SUCCESS. AT THAT TIME, HE ASKED FOR A NEW PERFORM HUMERAL STEM TO BE OPENED AND REVERSE INSERT. THE NEW PERFORM HUMERAL STEM AND INSERT WAS IMPLANTED WITHOUT ANY PROBLEMS. THIS WHOLE TROUBLESHOOTING PROCESS PROBABLY ADDED 20-30 MINUTES TO THE SURGICAL TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12787411 ↗
Other shoulder prosthesis, reverse configuration devices
- DJO SURGICAL 508-36-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration508 reports
- DJO SURGICAL 508-32-101Encore Medical, L.P. · shoulder prosthesis, reverse configuration433 reports
- DJO SURGICAL 508-32-103Encore Medical, L.P. · shoulder prosthesis, reverse configuration421 reports
- DJO SURGICAL 509-00-036Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall259 reports
- DJO SURGICAL 508-32-104Encore Medical, L.P. · shoulder prosthesis, reverse configuration1 recall203 reports
- DJO SURGICAL 509-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration233 reports
- DELTA XTEND 130760000Depuy (Ireland) · shoulder prosthesis, reverse configuration223 reports
- DJO SURGICAL 508-00-032Encore Medical, L.P. · shoulder prosthesis, reverse configuration122 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page