TORNIER PERFORM HUMERAL SYSTEMRecalls & reported problems across all models
Tornier, Inc. · 68 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one TORNIER PERFORM HUMERAL SYSTEM model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 190
- Deaths reported
- 1
- Injuries reported
- 170
- Recalls
- 1
We found 68 models of TORNIER PERFORM HUMERAL SYSTEM by Tornier, Inc. in the FDA's device registry (GUDID). Together they are named in 190 adverse event reports in the FDA's MAUDE database, received between Nov 10, 2021 and Sep 28, 2026 — classified as 1 death, 170 injuries, 19 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (48 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
TORNIER PERFORM HUMERAL SYSTEM recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Aug 29, 2024Class IIOpen, Classified
Devices from one lot of Perform" Reversed Inserts may contain a locking ring that was assembled upside down which would prevent the humeral insert from securely seating in the stem or spacer.
Z-3309-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported TORNIER PERFORM HUMERAL SYSTEM problems
Device problems
- Adverse Event Without Identified Device or Use Problem109
- Device Dislodged or Dislocated48
- Unstable11
- Device Difficult to Setup or Prepare7
- Fracture5
- Migration4
- Osseointegration Problem4
- Detachment of Device or Device Component3
Patient problems
- Joint Dislocation35
- Pain27
- Implant Pain22
- Insufficient Information21
- No Clinical Signs, Symptoms or Conditions18
- Unspecified Infection15
- Joint Laxity14
- Loss of Range of Motion14
TORNIER PERFORM HUMERAL SYSTEM models
68 models, most reported first. Each model has its own page with its reports, recalls and device details.
- TORNIER PERFORM™ HUMERAL SYSTEM DWP2393shoulder prosthesis, reverse configuration1 recall2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWX3SSPROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED14 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWX2PSprosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented12 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP1360shoulder prosthesis, reverse configuration10 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWX3PSPROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED10 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP2390shoulder prosthesis, reverse configuration9 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP2360shoulder prosthesis, reverse configuration6 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWX934shoulder prosthesis, reverse configuration6 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWR4426shoulder prosthesis, reverse configuration5 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWT110PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED5 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWX2SSPROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED5 reports
- Tornier Perform Humeral System ARS1022306PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED5 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWN4316PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED4 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP2363shoulder prosthesis, reverse configuration4 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP2420shoulder prosthesis, reverse configuration4 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWR2390shoulder prosthesis, reverse configuration4 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWT100PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED4 reports
- TORNIER PERFORM™ HUMERAL SYSTEM MWM035PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED3 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP1393shoulder prosthesis, reverse configuration3 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP4393shoulder prosthesis, reverse configuration3 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWX1PLprosthesis, shoulder, semi-constrained, metal/polymer + additive, cemented3 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWX4PSPROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED1 death3 reports
- Tornier Perform Humeral System PFX08130SPROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED3 reports
- Tornier Perform Humeral System PFX09130PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED3 reports
- TORNIER PERFORM™ HUMERAL SYSTEM MWX135PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWN4115PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWN5019PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP1363shoulder prosthesis, reverse configuration2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP1366shoulder prosthesis, reverse configuration2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP1390shoulder prosthesis, reverse configuration2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP2396shoulder prosthesis, reverse configuration2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP2423shoulder prosthesis, reverse configuration2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP3390shoulder prosthesis, reverse configuration2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWP4396shoulder prosthesis, reverse configuration2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWR4420shoulder prosthesis, reverse configuration2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWX1SSPROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWX2PLPROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWX4PLPROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM DWX4SSPROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED2 reports
- Tornier Perform Humeral System PFX10130PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED2 reports
- TORNIER PERFORM™ HUMERAL SYSTEM MWX4SSPROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED1 report
- TORNIER PERFORM™ HUMERAL SYSTEM MWM001PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWN4314PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWN4816PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWN4818PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWP1396shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWP2366shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWP2426shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWP3360shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWP4390shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWP4426shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWR1363shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWR1390shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWR1393shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWR1396shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWR2420shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWR2426shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWR3360shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWR4393shoulder prosthesis, reverse configuration1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWT4617PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 report
- TORNIER PERFORM™ HUMERAL SYSTEM DWX3PLPROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 report
- Tornier Perform Humeral System IFX801PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED1 report
- Tornier Perform Humeral System PFX09130SPROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED1 report
- Tornier Perform Humeral System PFX11130PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED1 report
- Tornier Perform Humeral System PFX14130PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 report
- Tornier Perform Humeral System PFX15130shoulder prosthesis, reverse configuration1 report
- Tornier Perform Humeral System PFX00120PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 report
- 1 report
Recent TORNIER PERFORM HUMERAL SYSTEM adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 28, 2026Loss of Osseointegration
Patient problems: Inadequate Osseointegration; Joint Dislocation
MAUDE report 50922232 ↗InjurySep 28, 2026Device Dislodged or Dislocated
Patient problems: Joint Dislocation
MAUDE report 50922228 ↗MalfunctionSep 1, 2026Use of Device Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50540926 ↗InjuryJul 27, 2026Adverse Event Without Identified Device or Use Problem
All implants were removed due to suspected infection.
Patient problems: Post Operative Wound Infection
MAUDE report 50042216 ↗InjuryJul 7, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE DUE TO SUSPECTED INFECTION. DURING THE PROCEDURE, THE IMPLANT WAS CLEANED, AND THE INSERT WAS REMOVED AND REPLACED.
Patient problems: Post Operative Wound Infection
MAUDE report 25736201 ↗MalfunctionJun 22, 2026Fracture
IT WAS REPORTED BY THE CUSTOMER THAT ONE OF THE INSTRUMENT'S PEGS BROKE. NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25579190 ↗InjuryJun 19, 2026Loosening of Implant Not Related to Bone-Ingrowth
THE PATIENT SUSTAINED A PROXIMAL HUMERUS FRACTURE THAT WAS TREATED WITH TUBEROSITY REPAIR AND HEMIARTHROPLASTY. AT THE TIME OF THE PROCEDURE, THE ROTATOR CUFF WAS CONFIRMED TO BE INTACT. FOLLOWING SURGERY, THE PATIENT WAS REPORTEDLY NON-COMPLIANT WITH SLING USE AND NON-WEIGHT-BEARING RESTRICTIONS, RESULTING IN FAILURE OF THE FRACTURE REPAIR. THE PATIENT DID NOT RETURN FOR FOLLOW-UP UNTIL A LATER DATE, AT WHICH TIME CONCERNS REGARDING THE OUTCOME OF THE REPAIR WERE IDENTIFIED. FOLLOWING POSITIVE CULTURE RESULTS, THE PATIENT REQUESTED REMOVAL OF THE HEMIARTHROPLASTY AND PLACEMENT OF AN ANTIBIOTIC SPACER. THE SURGEON EXPLAINED THE ANTICIPATED FUNCTIONAL LIMITATIONS ASSOCIATED WITH REMOVAL OF THE IMPLANT; HOWEVER, THE PATIENT ELECTED TO PROCEED WITH EXPLANTATION. THE HEMIARTHROPLASTY WAS REMOVED AND REPLACED WITH AN ANTIBIOTIC SPACER.
Patient problems: Loss of Range of Motion; Post Operative Wound Infection
MAUDE report 25563777 ↗InjuryJun 19, 2026Loosening of Implant Not Related to Bone-Ingrowth
THE PATIENT SUSTAINED A PROXIMAL HUMERUS FRACTURE THAT WAS TREATED WITH TUBEROSITY REPAIR AND HEMIARTHROPLASTY. AT THE TIME OF THE PROCEDURE, THE ROTATOR CUFF WAS CONFIRMED TO BE INTACT. FOLLOWING SURGERY, THE PATIENT WAS REPORTEDLY NON-COMPLIANT WITH SLING USE AND NON-WEIGHT-BEARING RESTRICTIONS, RESULTING IN FAILURE OF THE FRACTURE REPAIR. THE PATIENT DID NOT RETURN FOR FOLLOW-UP UNTIL A LATER DATE, AT WHICH TIME CONCERNS REGARDING THE OUTCOME OF THE REPAIR WERE IDENTIFIED. FOLLOWING POSITIVE CULTURE RESULTS, THE PATIENT REQUESTED REMOVAL OF THE HEMIARTHROPLASTY AND PLACEMENT OF AN ANTIBIOTIC SPACER. THE SURGEON EXPLAINED THE ANTICIPATED FUNCTIONAL LIMITATIONS ASSOCIATED WITH REMOVAL OF THE IMPLANT; HOWEVER, THE PATIENT ELECTED TO PROCEED WITH EXPLANTATION. THE HEMIARTHROPLASTY WAS REMOVED AND REPLACED WITH AN ANTIBIOTIC SPACER.
Patient problems: Loss of Range of Motion; Post Operative Wound Infection
MAUDE report 25563263 ↗InjuryMay 29, 2026Adverse Event Without Identified Device or Use Problem
AS REPORTED: "BECAUSE THE PATIENT WAS EXPERIENCING PAIN, A REVISION SURGERY WAS PERFORMED TO DETERMINE IF THE IMPLANT WAS LOOSE OR NEEDED RESIZING. A SLIGHT FOREIGN BODY REACTION SUGGESTIVE OF INFECTION WAS OBSERVED."
Patient problems: Pain; Unspecified Infection
MAUDE report 25337042 ↗InjuryMay 19, 2026Device Dislodged or Dislocated
THE PROCEDURE TOOK PLACE DUE TO CHRONIC DISLOCATION. ONCE THE SURGEON GOT EXPOSURE, THE TRAY AND POLY WERE REMOVED AND THE POLY HAD MEDIAL WEAR.
Patient problems: Insufficient Information
MAUDE report 25217218 ↗MalfunctionMay 6, 2026Fracture
IT WAS REPORTED THAT LATERALITY IS MISSING FROM HANDLE. THE LOCKING MECHANISM IS BROKEN. THE PHYSICIAN ATTEMPTED TO FIX IT BUT IT DIDN'T WORK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25101773 ↗MalfunctionMay 1, 2026Fracture
DURING USE, THE SLAPHAMMER FRACTURED WHILE ENGAGED WITH A SIZE 4 TRIAL STEM. ADDITIONALLY, TWO SIZE 2 SIZER DISCS WERE NOTED TO BE BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25059572 ↗InjuryApr 28, 2026Detachment of Device or Device Component; Device Dislodged or Dislocated
IT WAS REPORTED THAT, THE PATIENT UNDERWENT A PLANNED RIGHT REVERSE TOTAL SHOULDER REPLACEMENT ON (B)(6) 2026. POST OPERATIVE DAY ONE, ROUTINE X RAY IMAGING IDENTIFIED A DISLOCATION OF THE GLENO HUMERAL JOINT REPLACEMENT, WITH ANTERIOR SUBLUXATION OF THE PROXIMAL HUMERUS. IMAGING ALSO DEMONSTRATED THAT THE METALLIC RING COMPONENT OF THE PROSTHESIS HAD DISSOCIATED FROM THE IMPLANT. THE PATIENT UNDERWENT REVISION SURGERY TO REMOVE AND REPLACE THESE COMPONENTS. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: Subluxation
MAUDE report 25016119 ↗MalfunctionApr 15, 2026Fracture
A REVISION SURGERY WAS PERFORMED DUE TO A BROKEN HUMERAL STEM WHICH WAS BROKEN AT THE END LEVEL OF THE SUPERIOR MOST LATERAL SCREW.
Patient problems: Insufficient Information
MAUDE report 24888828 ↗MalfunctionApr 7, 2026Material Erosion
IT WAS REPORTED THAT, WHEN INSERTING THE GUIDE PIN (3X75 MM) INTO THE IM CUTTING BLOCK 135° (MWX135), METAL ABRASION OCCURRED ON THE GUIDE PIN AT THE UPPER 0° HOLE FOR THE GUIDE PINS FOR RESECTION OF THE RIGHT SHOULDER. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24801419 ↗
Other Tornier, Inc. products
- Tornier Perform Reversed Glenoid41 models · 515 reports · 1 recall
- Tornier Perform40 models · 290 reports · 1 recall
- Tornier HRS40 models · 176 reports · 1 recall
- Tornier Perform Reversed Augmented Glenoid8 models · 121 reports · 1 recall
- AEQUALIS FLEX REVIVE24 models · 70 reports · 2 recalls
- Latitude - Elbow35 models · 147 reports
- Tornier Latitude7 models · 14 reports · 1 recall
- Latitude EV24 models · 56 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page