Skip to content

TORNIER PERFORM™ HUMERAL SYSTEM DWR1396Recalls & reported problems

Tornier, Inc. · Class 2 · Implant

All models of TORNIER PERFORM HUMERAL SYSTEM (68)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00846832084807), received between Mar 11, 2022 and Mar 11, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • Fall1
  • Implant Pain1
  • Loss of Range of Motion1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 11, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT WAS HAPPY WITH HIS PROGRESS AND PERFORMING ACTIVITIES OF DAILY LIVING WITHOUT COMPLAINT POST RSA IN (B)(6) OF 2020. THE PATIENT HAD A 39 STANDARD GLENOSPHERE ON A 29 +3 BASEPLATE WITH A LONG POST. ON THE HUMERAL SIDE HE HAD A 2+ STEM AND A 10 DEGREE ½ 39 +6 POLY. THE PATIENT HAD A FALL. HE HAD PAIN AND LIMITED FUNCTION AFTER THE FALL. THE PATIENT WAS SEEN IN THE ER AND APPEARED TO HAVE A POLY DISSOCIATION. REVISION SURGERY WAS SCHEDULED FOR LATER THAT DAY ((B)(6) 2022). REVISION SURGERY ((B)(6) 2022): THE POLY WAS COMPLETELY OUT OF THE STEM. THE RING WAS RETAINED WITHIN THE STEM. BOTH THE POLY AND LOCKING RING WERE REMOVED. A SPACER WITH +0 POLY WAS TRIALED WITH GOOD STABILITY AND MOTION. THIS WAS IMPLANTED IN THE PATIENT.

    Patient problems: Fall; Implant Pain; Loss of Range of Motion

    MAUDE report 13734685 ↗

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page