Skip to content

Tornier Perform® Reversed Glenoid DWJ342Recalls & reported problems

Tornier, Inc. · Peripheral Screw · Class 2 · Implant

All models of Tornier Perform Reversed Glenoid (41)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00846832088904), received between Sep 19, 2022 and Jul 12, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Fracture3
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 12, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT AFTER THE BASEPLATE WAS INSERTED, WHILE TIGHTENING THE PERIPHERAL SCREWS, ONE OF THE 42MM SCREWS SNAPPED ONE CENTIMETER BELOW THE HEAD. THE SURGEON DETERMINED THAT THE FIXATION WAS STABLE ENOUGH WITH THE REMAINING THREE PERIPHERAL SCREWS AND THE CENTRAL SCREW, SO THE BROKEN 42MM SCREW WAS LEFT IN PLACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19735564 ↗
  • MalfunctionJan 25, 2023Break; Fracture; Mechanical Jam; Break; Fracture; Mechanical Jam

    IT WAS REPORTED THAT 3 PERIPHERAL SCREWS BROKE OFF AT THE NECK OF THE SCREW WHILE IMPLANTING. ALL 3 INCIDENT HAVE BEEN THE EXACT SAME SCENARIO IN ALL 3 CASES THE SURGEON DRILLED THE PERIPHERAL SCREW HOLE AND THEN MEASURED FOR SCREW DEPTH AND STARTED TO IMPLANT CORRESPONDING SCREW. ONCE THE SCREW WAS ABOUT 90% IMPLANTED IT WOULD STOPPED TURNING. THE SURGEON IN EACH SCENARIO THEN ATTEMPTED TO BACK OUT THE SCREW AND IT WOULD NOT BACK OUT OR ADVANCE ANY MORE. THE SURGEON THEN CONTINUED TRYING TO BACK OUT THE SCREW UNTIL IT FINALLY SNAPPED 3-4 THREADS DOWN THE SCREW. THE SURGEON WENT ABOUT A NORMAL TECHNIQUE TO COMPLETE THE SURGERY. ALL THE DEBRIS WAS SUCCESSFULLY REMOVED FROM THE FIELD/PATIENT. PRODUCTS WILL NOT BE RETURNING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16240510 ↗
  • MalfunctionSep 19, 2022Break; Fracture; Break; Fracture

    IT WAS REPORTED A PATIENT WENT INTO SURGERY FOR A LEFT REVERSE TOTAL SHOULDER REPLACEMENT. WHILE THE SURGEON WAS IMPLANTING THE BASEPLATE, ONE SCREW BROKE DURING THE FASTENING PROCESS. BASEPLATE SECURED WITH REMAINING SCREW AND OTHER SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15451280 ↗

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page