Tornier Perform® Reversed Glenoid DWJ342Recalls & reported problems
Tornier, Inc. · Peripheral Screw · Class 2 · Implant
All models of Tornier Perform Reversed Glenoid (41)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00846832088904), received between Sep 19, 2022 and Jul 12, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Fracture3
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJul 12, 2024Break; Fracture; Break; Fracture
IT WAS REPORTED THAT AFTER THE BASEPLATE WAS INSERTED, WHILE TIGHTENING THE PERIPHERAL SCREWS, ONE OF THE 42MM SCREWS SNAPPED ONE CENTIMETER BELOW THE HEAD. THE SURGEON DETERMINED THAT THE FIXATION WAS STABLE ENOUGH WITH THE REMAINING THREE PERIPHERAL SCREWS AND THE CENTRAL SCREW, SO THE BROKEN 42MM SCREW WAS LEFT IN PLACE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19735564 ↗MalfunctionJan 25, 2023Break; Fracture; Mechanical Jam; Break; Fracture; Mechanical Jam
IT WAS REPORTED THAT 3 PERIPHERAL SCREWS BROKE OFF AT THE NECK OF THE SCREW WHILE IMPLANTING. ALL 3 INCIDENT HAVE BEEN THE EXACT SAME SCENARIO IN ALL 3 CASES THE SURGEON DRILLED THE PERIPHERAL SCREW HOLE AND THEN MEASURED FOR SCREW DEPTH AND STARTED TO IMPLANT CORRESPONDING SCREW. ONCE THE SCREW WAS ABOUT 90% IMPLANTED IT WOULD STOPPED TURNING. THE SURGEON IN EACH SCENARIO THEN ATTEMPTED TO BACK OUT THE SCREW AND IT WOULD NOT BACK OUT OR ADVANCE ANY MORE. THE SURGEON THEN CONTINUED TRYING TO BACK OUT THE SCREW UNTIL IT FINALLY SNAPPED 3-4 THREADS DOWN THE SCREW. THE SURGEON WENT ABOUT A NORMAL TECHNIQUE TO COMPLETE THE SURGERY. ALL THE DEBRIS WAS SUCCESSFULLY REMOVED FROM THE FIELD/PATIENT. PRODUCTS WILL NOT BE RETURNING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16240510 ↗MalfunctionSep 19, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED A PATIENT WENT INTO SURGERY FOR A LEFT REVERSE TOTAL SHOULDER REPLACEMENT. WHILE THE SURGEON WAS IMPLANTING THE BASEPLATE, ONE SCREW BROKE DURING THE FASTENING PROCESS. BASEPLATE SECURED WITH REMAINING SCREW AND OTHER SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15451280 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page