Tornier Perform® Reversed Glenoid DWJ338Recalls & reported problems
Tornier, Inc. · Peripheral Screw · Class 2 · Implant
All models of Tornier Perform Reversed Glenoid (41)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00846832089307), received between Apr 13, 2023 and Mar 2, 2026 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- Break2
- Fracture2
- Migration1
Patient problems
- Pain2
- Unspecified Infection2
- Failure of Implant1
- Fluid Discharge1
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
- Post Operative Wound Infection1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryMar 2, 2026Adverse Event Without Identified Device or Use Problem
THE PATIENT UNDERWENT A REVISION SURGERY. PATIENT HAD SHOWN SIGNS OF INFECTION AND WAS DRAINING FROM THE HEALED INCISION SITE. NO INSTABILITY WAS REPORTED OR OBSERVED. CULTURES WERE PERFORMED INTRAOPERATIVELY. THE PATIENTS 3YR OLD WOUND WAS DRAINING FOR THE LAST 3-4 WEEKS AND THE DOCTOR DECIDED TO REMOVE ALL THE IMPLANTS AND IMPLANT AN ANTIBIOTIC SPACER.
Patient problems: Fluid Discharge; Post Operative Wound Infection
MAUDE report 24482499 ↗InjuryJul 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PATIENT UNDERWENT REVISION SURGERY DUE TO ALLEGED PAIN AND SUPERIOR MIGRATION OF THE GLENOID IMPLANT.
Patient problems: Pain
MAUDE report 22375914 ↗InjuryJul 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PATIENT UNDERWENT REVISION SURGERY DUE TO ALLEGED PAIN AND SUPERIOR MIGRATION OF THE GLENOID IMPLANT.
Patient problems: Pain
MAUDE report 22375912 ↗InjuryMay 16, 2024Break; Fracture; Migration; Break; Fracture; Migration
THE PRIMARY REVERSE PROCEDURE EMPLOYED TORNIER IMPLANTS, INCLUDING THE SHOULDER ID BASEPLATE AND PERFORM HUMERAL. THE OBSERVED BASEPLATE FAILURE IS ATTRIBUTED TO LIKELY OVER-LATERALIZATION, STEMMING FROM THE THICKNESS OF THE SHOULDER ID BASEPLATE AND A LATERALIZED GLENOSPHERE. THE SCREW ALSO HAD A BREAKAGE.
Patient problems: Failure of Implant
MAUDE report 19328764 ↗InjuryFeb 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REVISION SURGERY WAS CONDUCTED IN RESPONSE TO AN INFECTION. AN UPDATE PROVIDED INDICATES A LACK OF INFORMATION REGARDING THE LOT NUMBERS FOR THE SCREWS USED. HOWEVER, IT WAS MENTIONED THAT THE INFECTION APPEARED TO BE LOCALIZED, AND THE PROCEDURE INVOLVED WASHING OUT AND EXCHANGING PARTS.
Patient problems: Unspecified Infection
MAUDE report 18784091 ↗InjuryJun 29, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED WITH SUPERFICIAL INFECTION. THE PHYSICIAN TREATED THE INFECTION WITH IRRIGATION AND DEBRIDEMENT WITH SALINE AND VANCOMYCIN POWDER. ALL THE HARDWARE WAS LEFT IN THE PATIENT AS THE DOCTOR DEEMED THE INFECTION WAS NOT DEEP ENOUGH TO HAVE BEEN A RESULT OF THE IMPLANTS. THE PATIENT WAS ADMINISTERED AND PRESCRIBED ANTIBIOTICS.
Patient problems: Unspecified Infection
MAUDE report 17233564 ↗MalfunctionApr 13, 2023Break; Fracture; Break; Fracture
IT WAS REPORTED THAT A SCREW BROKE WHILE TRYING TO BACK IT OUT. IT WAS FURTHER REPORTED THAT BROKEN HALF OF THE SCREW WAS LEFT IMPLANTED. THE REMAINING HALF WAS REMOVED AND WILL BE RETURNED FOR INVESTIGATION.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 16738475 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page