Tornier Perform® Reversed Glenoid DWJ322Recalls & reported problems
Tornier, Inc. · Peripheral Screw · Class 2 · Implant
All models of Tornier Perform Reversed Glenoid (41)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00846832090051), received between Feb 26, 2024 and Oct 21, 2025 — classified as 7 injuries.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem6
- Break1
- Fracture1
- Migration1
Patient problems
- Unspecified Infection4
- Bone Fracture(s)1
- Failure of Implant1
- Fall1
- Post Operative Wound Infection1
- Rash1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryOct 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED USING THE BLUEPRINT PLANNING FEATURE. THE PATIENT HAD PREVIOUSLY UNDERGONE A PRIMARY SHOULDER PROCEDURE INVOLVING STRYKER/TORNIER IMPLANTS. FOLLOWING THE INITIAL SURGERY, THE PATIENT DEVELOPED AN INFECTION, AND ALL IMPLANTS WERE EXPLANTED WITH PLACEMENT OF AN ANTIBIOTIC SPACER. THE REVISION PROCEDURE INVOLVED REMOVAL OF THE SPACER AND CONVERSION TO A REVERSE SHOULDER ARTHROPLASTY.
Patient problems: Post Operative Wound Infection
MAUDE report 23347194 ↗InjuryJan 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT FELL RESULTING IN A PERIPROSTHETIC FRACTURE. ALL IMPLANTED COMPONENTS WERE REMOVED, AND A REVERSE PROSTHESIS WAS PLACED. NO IMAGES WERE PROVIDED. THE EXPLANTED IMPLANTS WERE NOT AVAILABLE FOR EVALUATION PER HOSPITAL POLICY REQUIRING DISPOSAL OF EXPLANTED ITEMS.
Patient problems: Bone Fracture(s); Fall
MAUDE report 21242131 ↗InjurySep 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT A ONE STAGE REVISION DUE TO AN ACUTE INFECTION.
Patient problems: Unspecified Infection
MAUDE report 20327396 ↗InjurySep 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT A ONE STAGE REVISION DUE TO AN ACUTE INFECTION.
Patient problems: Unspecified Infection
MAUDE report 20327242 ↗InjuryMay 16, 2024Break; Fracture; Migration; Break; Fracture; Migration
THE PRIMARY REVERSE PROCEDURE EMPLOYED TORNIER IMPLANTS, INCLUDING THE SHOULDER ID BASEPLATE AND PERFORM HUMERAL. THE OBSERVED BASEPLATE FAILURE IS ATTRIBUTED TO LIKELY OVER-LATERALIZATION, STEMMING FROM THE THICKNESS OF THE SHOULDER ID BASEPLATE AND A LATERALIZED GLENOSPHERE. THERE WAS ALSO A SCREW BREAKAGE.
Patient problems: Failure of Implant
MAUDE report 19328765 ↗InjuryApr 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING A PHONE CALL, A CASE OF REVISION SURGERY FOR A JOINT INFECTION RELATED TO INSTRUMENT BREAKAGE WAS DISCUSSED. THE BROKEN INSTRUMENT, IDENTIFIED AS A SCREWDRIVER, FRACTURED DURING A REVERSE SHOULDER REVISION SURGERY. THE PROCEDURE INVOLVED SUCCESSFULLY IMPLANTING A DEPUY SPACER. THE INFECTION WAS DETECTED THROUGH A SUPERFICIAL RASH NEAR THE INCISION SITE. MICROORGANISMS WERE NOT IDENTIFIED, AND IT'S UNCERTAIN IF THE PATIENT HAD A PRE-EXISTING INFECTION. NO STERILE IMPLANTS WERE OPENED DURING THE SURGERY, AND NO RE-STERILIZATION OF THE DEVICE WAS PERFORMED BEFORE THE SURGERY.
Patient problems: Rash; Unspecified Infection
MAUDE report 19038420 ↗InjuryFeb 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE REVISION SURGERY WAS CONDUCTED IN RESPONSE TO AN INFECTION. AN UPDATE PROVIDED INDICATES A LACK OF INFORMATION REGARDING THE LOT NUMBERS FOR THE SCREWS USED. HOWEVER, IT WAS MENTIONED THAT THE INFECTION APPEARED TO BE LOCALIZED, AND THE PROCEDURE INVOLVED WASHING OUT AND EXCHANGING PARTS.
Patient problems: Unspecified Infection
MAUDE report 18778202 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page