Tornier Perform® Reversed Glenoid DWJ742Recalls & reported problems
Tornier, Inc. · Peripheral Screw · Class 2 · Implant
All models of Tornier Perform Reversed Glenoid (41)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00846832091287), received between Mar 8, 2022 and Sep 16, 2026 — classified as 6 injuries.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Break1
- Fracture1
- Loss of Osseointegration1
Patient problems
- Unspecified Infection4
- Failure of Implant1
- Inadequate Osseointegration1
- Pain1
- Swelling/ Edema1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem
As reported: "Patient has re-presented to surgeon, following revision shoulder arthroplasty in (b)(6)2026. Significant swelling and redness over wound, pain on movement. Wound re-opened, and surgeon concluded that the prosthetic joint was now infected. Decided to remove all prosthesis, as stage 1 of a 2-stage revision. Glenoid component was found to be very loose, easily removed as 1 x construct (revised at previous revision). Humeral component (not revised at previous revision) was found to be extremely well fixed. Humeral poly insert removed - significant wear noted (only been insitu since (b)(6) 2026 revision). Proximal humeral vertical osteotomy was required to loosen and to break the bone / implant ongrowth. Stem eventually removed; osteotomy cabled back together with Nice Loops. Multiple tissue specimens collected; debridement; copious lavage attended; no antibiotic spacer implanted - gleno-humeral joint left dis-articulated.Plan - antibiotic regime until clinical markers indicate no infection, then the 2nd stage revision will be undertaken, using a "Custom Glenoid", to be manufactured by a competitor company."
Patient problems: Pain; Swelling/ Edema; Unspecified Infection
MAUDE report 50791182 ↗InjuryOct 28, 2025Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT, THE REVISION SURGERY WAS PERFORMED AND THE ASCEND FLEX/REVERSED II IMPLANTS WERE REVISED TO THE PERFORM REVERSE GLENOID IMPLANTS AS THE GLENOID IMPLANTS HAD MIGRATED SUPERIORLY. REVISION SURGERY WAS SUCCESSFULLY COMPLETED. IT WAS FURTHER REPORTED THAT, ALL GLENOID IMPLANTS WERE REMOVED (SUSPECTS), THE HUMERAL TRAY AND LINER WAS REMOVED (CONCOMITANTS). HUMERAL STEM REMAINED STABLE AND SO SURGEON LEFT IN SITU. NEW TRAY AND LINER AND GLENOID SYSTEM INSERTED SUCCESSFULLY.
Patient problems: Inadequate Osseointegration
MAUDE report 23405470 ↗InjuryAug 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN (B)(6) 2012 ¿ (B)(6) 2022. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2020 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. REVISION PROCEDURE TYPE: STAGE 1 OF 2 STAGE REVISION.
Patient problems: Unspecified Infection
MAUDE report 19919047 ↗InjuryMay 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THERE WAS A REVISION SURGERY CAUSED BY AN INFECTION. ADDITIONALLY, ALL TEN IMPLANTS WERE REMOVED. THE NURSE STILL HAD THE CENTRAL SCREW WITHIN THE BASEPLATE AND THE REVERSED INSERT WAS UNATTACHED FROM THE REVERSED TRAY.
Patient problems: Unspecified Infection
MAUDE report 16989719 ↗InjuryJul 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION SURGERY DUE TO AN INFECTION 12 MONTHS POST-OP. THE METAL DEVICES WERE REMOVED AND A COMPETITOR'S CEMENT SPACER WAS INSERTED.
Patient problems: Unspecified Infection
MAUDE report 15043773 ↗InjuryMar 8, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION DUE TO GLENOID BASEPLATE LOOSENING AND A SNAPPED SUPERIOR SCREW. DURING THE REVISION IT WAS CONFIRMED THAT THE BASEPLATE WAS LOOSE IN THE PATIENT'S GLENOID. THERE WAS ALSO EXCESSIVE WEAR ON THE HUMERAL POLY INSERT. THE HUMERAL STEM WAS WELL FIXED.
Patient problems: Failure of Implant
MAUDE report 13691014 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page