The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00846832091287), received between Mar 8, 2022 and Sep 16, 2026 — classified as 6 injuries.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Break1
  • Fracture1
  • Loss of Osseointegration1

Patient problems

  • Unspecified Infection4
  • Failure of Implant1
  • Inadequate Osseointegration1
  • Pain1
  • Swelling/ Edema1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem

    As reported: "Patient has re-presented to surgeon, following revision shoulder arthroplasty in (b)(6)2026. Significant swelling and redness over wound, pain on movement. Wound re-opened, and surgeon concluded that the prosthetic joint was now infected. Decided to remove all prosthesis, as stage 1 of a 2-stage revision. Glenoid component was found to be very loose, easily removed as 1 x construct (revised at previous revision). Humeral component (not revised at previous revision) was found to be extremely well fixed. Humeral poly insert removed - significant wear noted (only been insitu since (b)(6) 2026 revision). Proximal humeral vertical osteotomy was required to loosen and to break the bone / implant ongrowth. Stem eventually removed; osteotomy cabled back together with Nice Loops. Multiple tissue specimens collected; debridement; copious lavage attended; no antibiotic spacer implanted - gleno-humeral joint left dis-articulated.Plan - antibiotic regime until clinical markers indicate no infection, then the 2nd stage revision will be undertaken, using a "Custom Glenoid", to be manufactured by a competitor company."

    Patient problems: Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 50791182 ↗
  • InjuryOct 28, 2025Loss of Osseointegration; Loss of Osseointegration

    IT WAS REPORTED THAT, THE REVISION SURGERY WAS PERFORMED AND THE ASCEND FLEX/REVERSED II IMPLANTS WERE REVISED TO THE PERFORM REVERSE GLENOID IMPLANTS AS THE GLENOID IMPLANTS HAD MIGRATED SUPERIORLY. REVISION SURGERY WAS SUCCESSFULLY COMPLETED. IT WAS FURTHER REPORTED THAT, ALL GLENOID IMPLANTS WERE REMOVED (SUSPECTS), THE HUMERAL TRAY AND LINER WAS REMOVED (CONCOMITANTS). HUMERAL STEM REMAINED STABLE AND SO SURGEON LEFT IN SITU. NEW TRAY AND LINER AND GLENOID SYSTEM INSERTED SUCCESSFULLY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23405470 ↗
  • InjuryAug 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE MANUFACTURER RECEIVED A REPORT FROM THE NATIONAL JOINT REGISTRY THAT CONTAINS UNPUBLISHED COLLECTED DATA ON THE USAGE AND THE OUTCOMES WITH THE TORNIER SHOULDER SYSTEM. THE REPORT DETAILS ANALYSIS PROVIDED FOR PROCEDURES PERFORMED BETWEEN (B)(6) 2012 ¿ (B)(6) 2022. DURING THE REVIEW OF THE REPORT, IT WAS IDENTIFIED THAT ON (B)(6) 2020 A PATIENT REQUIRED REVISION SURGERY DUE TO INFECTION, WHICH WAS NOT PREVIOUSLY REPORTED TO THE MANUFACTURER. REVISION PROCEDURE TYPE: STAGE 1 OF 2 STAGE REVISION.

    Patient problems: Unspecified Infection

    MAUDE report 19919047 ↗
  • InjuryMay 23, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THERE WAS A REVISION SURGERY CAUSED BY AN INFECTION. ADDITIONALLY, ALL TEN IMPLANTS WERE REMOVED. THE NURSE STILL HAD THE CENTRAL SCREW WITHIN THE BASEPLATE AND THE REVERSED INSERT WAS UNATTACHED FROM THE REVERSED TRAY.

    Patient problems: Unspecified Infection

    MAUDE report 16989719 ↗
  • InjuryJul 18, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION SURGERY DUE TO AN INFECTION 12 MONTHS POST-OP. THE METAL DEVICES WERE REMOVED AND A COMPETITOR'S CEMENT SPACER WAS INSERTED.

    Patient problems: Unspecified Infection

    MAUDE report 15043773 ↗
  • InjuryMar 8, 2022Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION DUE TO GLENOID BASEPLATE LOOSENING AND A SNAPPED SUPERIOR SCREW. DURING THE REVISION IT WAS CONFIRMED THAT THE BASEPLATE WAS LOOSE IN THE PATIENT'S GLENOID. THERE WAS ALSO EXCESSIVE WEAR ON THE HUMERAL POLY INSERT. THE HUMERAL STEM WAS WELL FIXED.

    Patient problems: Failure of Implant

    MAUDE report 13691014 ↗

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices