ConcertoTM PV-16-40-HELIX (UDI 00847536017078)Recalls & reported problems
Micro Therapeutics, Inc. · PV-16-40-HELIX CONCERTO PGLA COIL · Class 2 · Implant
All models of ConcertoTM (70)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00847536017078), received between Apr 11, 2019 and Dec 13, 2023 — classified as 1 death, 3 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Separation Failure3
- Adverse Event Without Identified Device or Use Problem1
- Material Deformation1
- No Apparent Adverse Event1
Patient problems
- No Known Impact Or Consequence To Patient2
- Myocardial Infarction1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
DeathDec 13, 2023Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event
ADDITIONAL INFORMATION RECEIVED FROM THE COMMUNICATING AUTHOR REPORTED THE PATIENT'S DEATH WAS NOT ASSOCIATED WITH THE CONCERTO COILS. IMPLANTED CONCERTO COILS: INCLUDED HELIX EV3 CONCERTO PV 1640 LOT A366207; HELIX EV3 CONCERTO PV 1640 LOT A357326; HELIX EV3 CONCERTO PV 1640 LOT A364247; HELIX EV3 CONCERTO PV 1640 LOT A364247; HELIX EV3 CONCERTO PV 1430 LOT A366201; HELIX EV3 CONCERTO PV 1430 LOT A374709; HELIX EV3 CONCERTO PV 1430 LOT A 378790; HELIX EV3 CONCERTO PV 1640 LOT A 366207 AS THERE WAS NO REPORTED MALFUNCTION AND THE ADVERSE EVENT IS NOT ASSOCIATED WITH THE MEDTRONIC DEVICE(S), THIS EVENT IS NO LONGER A REPORTABLE EVENT. MDR DECISION CORRECTED TO NOT REPORTABLE. NO ADDITIONAL SUPPLEMENTAL MDRS ARE REQUIRED UNLESS ADDITIONAL INFORMATION RECEIVED MAKES THE EVENT REPORTABLE.
Patient problems: Myocardial Infarction
MAUDE report 18322614 ↗MalfunctionJan 5, 2021Separation Failure; Material Deformation; Separation Failure; Material Deformation
MEDTRONIC RECEIVED A REPORT THAT THE DEVICE FAILED TO DETACH. THREE DETACHMENT ATTEMPTS WERE MADE WITH THE INSTANT DETACHER, AND A SECOND INSTANT DETACHER WAS NOT USED. IT WAS UNKNOWN IF MANUAL DETACHMENT ATTEMPTS WERE MADE. A COIL WAS USED IN THIS CASE TO AVOID AND OCCLUDE RETROGRADE BLOOD FLOW FROM THE RIGHT INTERNAL ILIAC ARTERY, AN OFF-LABEL USE. THE DEVICES WERE PREPARED AS INDICATED PER THE IFU. THERE WERE NO RELATED PATIENT SYMPTOMS. IT WAS NOTED THAT THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MODERATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11118782 ↗MalfunctionMay 12, 2020Separation Failure; Separation Failure
ADDITIONAL INFORMATION RECEIVED FROM A HEALTHCARE PROVIDER (HCP), VIA A MANUFACTURING REPRESENTATIVE (REP), INDICATED THE ANEURYSM MEASURED APPROXIMATELY 40X40 MM. THREE ATTEMPTS EACH WERE MADE WITH THE INSTANT DETACHER AND MANUAL METHODS. THERE WERE NO ISSUES ENCOUNTERED PRIOR TO THE NON-DETACHMENT, AND NO DAMAGE WAS OBSERVED TO THE PUSHWIRE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10048937 ↗MalfunctionApr 11, 2019Separation Failure; Separation Failure
MEDTRONIC RECEIVED REPORT THAT THE CONCERTO FAILED TO DETACH. THE PATIENT WAS UNDERGOING TREATMENT OF AN UNRUPTURED, AORTIC ANEURYSM. THE MAX DIAMETER WAS 35 MM WITH A 10MM NECK. THE ANATOMY WAS MODERATE IN TORTUOSITY AND THE BLOOD FLOW WAS NORMAL. THE REPORTED DEVICE AND ANY ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE IFU. DURING THE PROCEDURE, MULTIPLE COILS WERE PLACED. A CONCERTO COIL WAS THEN ATTEMPTED. AT DETACHMENT, THE COIL DID NOT DETACH AFTER TWO ATTEMPTS WITH AN INSTANT DETACHER AND ONE ATTEMPT USING THE MANUAL METHOD. THE COIL WAS REMOVED. THE PROCEDURE WAS CONTINUED. THERE WERE NO REPORTS OF PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8506587 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page