AxiumTM QC-1.5-2-HELIX (UDI 00847536017177)Recalls & reported problems
Micro Therapeutics, Inc. · QC-1.5-2-HELIX BARE AXIUM 1.5MM X 2CM · Class 2 · Implant
All models of AxiumTM (139)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847536017177), received between Mar 18, 2019 and Mar 18, 2019 — classified as 1 malfunction.
The most frequently reported specific device problem is “Separation Failure” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Separation Failure1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 18, 2019Separation Failure; Separation Failure
MEDTRONIC RECEIVED INFORMATION THAT THIS COIL WOULD NOT ABLE TO DETACH WITH USING INSTANT DETACHER OR WITH MANUAL METHOD (BREAKING HYPOTUBE METHOD; AN ALTERNATIVE METHOD PRESENTED IN THE INSTRUCTION FOR USE). THE PATIENT WAS UNDERGOING COILING TREATMENT FOR AN UNRUPTURED SACCULAR ANEURYSM LOCATED IN POSTERIOR COMMUNICATING ARTERY, MEASURING 2MMX8MM. THE VESSEL WAS MINIMAL TORTUOUS. THE BLOOD FLOW WAS NORMAL. PRIOR TO THE EVENT, THE MICROCATHETER REACHED TO THE LESION, AND THE FIRST SELECTED QC-1.5-2-HELIX COIL WITH ECHELON MICROCATHETER AND DELIVER THE COIL INTO THE TUMOR AND SUCCESSFULLY RELEASED. THE REPORTED COIL WAS SECOND COIL IN THE PROCEDURE, WHEN TRIED TO DETACH THE COIL, THE DID NOT DETACH WITH INSTANT DETACHER AFTER REPEATED ATTEMPT. THE PHYSICIAN ATTEMPTED TWO TIMES WITH MANUAL METHOD STILL COIL DID NOT DETACH. THE THIRD THE SPRING COIL QC-2-2-HELIX WAS REPLACED, AND IT CAN BE OPENED AND RELEASED SMOOTHLY. THE PROCEDURE COMPLETED SUCCESSFULLY. THE PATIENT'S CONDITION WAS STABLE AFTER SURGERY AND THE ANEURYSM DID NOT RUPTURE. REPORTED DEVICE AND ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE IFU. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8430761 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page