Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847536017269), received between Aug 26, 2019 and Aug 26, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    MEDTRONIC RECEIVED INFORMATION THAT AN AXIUM IMPLANT COIL COULD NOT BE PUSHED OUT OF THE INTRODUCER SHEATH DURING A PROCEDURE. DURING PREPARATION, THE INTRODUCER SHEATH HAD BEEN HELD VERTICALLY WHEN THE IMPLANT COIL WAS PULLED BACK INSIDE DURING HYDRATION. THE COIL WAS REPLACED TO CONTINUE THE PROCEDURE WITHOUT ANY IMPACT TO THE PATIENT. THE PATIENT WAS UNDERGOING EMBOLIZATION TREATMENT OF AN UNRUPTURED SACCULAR ANEURYSM MEASURING 3.6MM X 1.7MM LOCATED IN THE OPHTHALMIC SEGMENT OF THE LEFT INTERNAL CAROTID ARTERY (ICA). THE VASCULATURE WAS NORMAL IN TORTUOSITY AND THE BLOOD FLOW WAS NORMAL. THE AXIUM COIL WAS RETURNED FOR ANALYSIS AND FOUND TO BE DETACHED FROM THE PUSHWIRE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8932633 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.

  • Z-0873-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
  • Z-0874-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →