Skip to content

ConcertoTM PV-3-8-3D (UDI 00847536026339)Recalls & reported problems

Micro Therapeutics, Inc. · COIL PV-3-8-3D V01 CONCERTO 3D · Class 2 · Implant

All models of ConcertoTM (70)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00847536026339), received between May 15, 2019 and Oct 20, 2021 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break2
  • Physical Resistance/Sticking2
  • Premature Separation1
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 20, 2021Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking

    ADDITIONAL INFORMATION RECEIVED REPORTED THE COIL WAS STUCK IN THE PROTECTIVE SHEATH. NO DAMAGE WAS OBSERVED. DURING PREPARATION, THE INTRODUCER SHEATH WAS HELD VERTICALLY WHEN THE IMPLANT COIL WAS PULLED IN DURING HYDRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12665764 ↗
  • MalfunctionOct 20, 2021Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking

    ADDITIONAL INFORMATION RECEIVED REPORTED THE COIL WAS STUCK IN THE PROTECTIVE SHEATH. NO DAMAGE WAS OBSERVED. DURING PREPARATION, THE INTRODUCER SHEATH WAS HELD VERTICALLY WHEN THE IMPLANT COIL WAS PULLED IN DURING HYDRATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12665655 ↗
  • MalfunctionSep 10, 2019Premature Separation; Premature Separation

    MEDTRONIC RECEIVED REPORT THAT RESISTANCE WAS ENCOUNTERED WHEN THE CONCERTO COIL WAS ADVANCED THROUGH THE RENEGADE CATHETER. THE DEVICE DID NOT EXIT THE MICROCATHETER. A CONTINUOUS FLUSH WAS NOT USED DURING THE PROCEDURE. ANCILLARY DEVICES INCLUDE A RENEGADE STRAIGHT .021.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8982788 ↗
  • MalfunctionMay 15, 2019Separation Failure; Separation Failure

    MEDTRONIC RECEIVED INFORMATION THAT THE CONCERTO COIL COULD NOT BE DETACHED WITH THE INSTANT DETACHER. THE COIL WAS REPLACED WITH A COIL FROM ANOTHER MANUFACTURE AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER COMPLICATIONS. NO PATIENT INJURY WAS REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8612757 ↗

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page