ConcertoTM PV-10-30-3D (UDI 00847536026483)Recalls & reported problems
Micro Therapeutics, Inc. · COIL PV-10-30-3D CONCERTO 3D V01 · Class 2 · Implant
All models of ConcertoTM (70)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536026483), received between Mar 7, 2019 and Apr 20, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 20, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
MEDTRONIC RECEIVED A REPORT THAT THE CONCERTO COIL SELF DETACHED IN THE MICROCATHETER. THE COIL WAS REMOVED FROM THE PATIENT WITH A SNARE, AND NO ADDITIONAL SURGICAL OR MEDICAL INTERVENTION WAS REQUIRED. THERE WAS NO FRICTION OR DIFFICULTY DURING DELIVERY, AND NO REPOSITION OR DETACHMENT ATTEMPTS HAD BEEN MADE. THE COIL WAS REPLACED, AND THE PATIENT DID NOT EXPERIENCE ANY INJURY OR COMPLICATIONS. THE DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11697515 ↗MalfunctionMar 7, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
MEDTRONIC RECEIVED REPORT THAT THE .021 RENEGADE CATHETER WAS IN PLACE AND ONE PV 10-30 3D COIL WAS PLACED NICELY INSIDE SACK AND DETACHED WITH NO ISSUES (MANUAL DETACHMENT). A SECOND COIL, NV 10-30 WAS DEPLOYED. THE COIL WAS RETRACTED FOR RE-POSITION. AT THIS POINT SOMETHING WAS NOTED TO BE WRONG. THE COIL COULDN'T BE REALLY PUSHED OUT, SO IT WAS WITHDRAWN (THE ENTIRE COIL AND WIRE). THE WIRE WAS TOUGH TO WITHDRAW, BUT IT KEPT COMING, THEN AN UNFAMILIAR PART WAS FELT WHEN GRABBED, AS IF SOMETHING WAS UNRAVELING. THE REP STEPPED OUT OF THE ROOM. WHEN HE RETURNED A NEW COIL WAS BEING PLACED INSIDE THE ANEURYSM. IT WAS BELIEVED THAT THE FIRST COIL DID NOT DETACH OUTSIDE THE MICRO CATHETER. WHEN THE SECOND COIL WAS PUSHED THROUGH THE MICRO CATHETER, IT MUST OF INTER TWINED WITH THE FIRST COIL. MANUAL DETACHMENT WAS THE ONLY DETACHMENT METHOD USED. ONE ATTEMPT WAS MADE. THE DEVICES WERE REPORTED NOT TO HAVE BEEN USED PER THE INSTRUCTIONS FOR USE, AS A CONTINUOUS FLUSH WAS NOT USED. NO PATIENT INJURY OCCURRED. NO IMAGES ARE AVAILABLE. THE DEVICES WERE REPORTED TO HAVE BEEN DISCARDED AT THE SITE. THE UNRUPTURED, ANEURYSM WAS IN THE LEFT INTERNAL ILIAC. THE MAX AND NECK DIAMETER WERE BOTH
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8401369 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page