Skip to content

AxiumTM QC-4-12-HELIXRecalls & reported problems

Micro Therapeutics, Inc. · COIL QC-4-12-HELIX V04 AXIUM · Class 2 · Implant

All models of AxiumTM (139)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536029590), received between Apr 19, 2021 and Nov 12, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Apr 27, 2022Class IIOpen, Classified

    Due to incorrect size and configuration labeling of the detachable coil system.

    Z-1261-2022 · Root cause: Labeling Change Control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 12, 2021Separation Failure; Separation Failure

    MEDTRONIC RECEIVED INFORMATION REGARDING AN AXIUM COIL THAT FAILED TO DETACH. THAT PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A RUPTURED SACCULAR RIGHT POSTERIOR COMMUNICATING ANEURYSM. THE ANEURYSM MAX DIAMETER WAS 5.24MM AND THE NECK DIAMETER WAS 2.1MM. BLOOD FLOW WAS NORMAL. VESSEL TORTUOSITY WAS MODERATE. IT WAS REPORTED THAT THE AXIUM COIL AND ALL ACCESSORY DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THE COIL WAS DEPLOYED BUT COULD NOT BE DETACHED WITH THE INSTANT DETACHER (ID). 3 ATTEMPTS WERE MADE TO DETACH THE COIL WITH THE ID WITHOUT SUCCESS. 2 ATTEMPTS WERE THEN MADE AT MANUAL DETACHMENT WITH THE HYPOTUBE BUT THIS WAS ALSO UNSUCCESSFUL. THE COIL WAS REMOVED AND REPLACED WITH AN AXIUM 4X10 COIL THAT WAS DELIVERED AND DETACHED WITH THE SAME ID WITHOUT ISSUE. IT WAS NOTED THERE WAS NO PROBLEM WITH THE ID. THERE WAS NO HARM OR INJURY TO THE PATIENT. ANCILLARY DEVICES: CORDIS 6F SHEATH, CHAPERON 6F GUIDE CATHETER, EXCELSIOR SL-10 MICROCATHETER, SYNCHRO 14 GUIDEWIRE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12805120 ↗
  • MalfunctionApr 19, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    ADDITIONAL INFORMATION RECEIVED REPORTED NO DETACHMENT ATTEMPTS WERE MADE, AND THERE WAS NO KINK/DAMAGE OBSERVED TO THE PUSHWIRE. ANCILLARY DEVICES INCLUDE AN ECHELON 10 MICROCATHETER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11688358 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.

  • Z-0874-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
  • Z-0873-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page