AxiumTM QC-7-30-HELIXRecalls & reported problems
Micro Therapeutics, Inc. · COIL QC-7-30-HELIX V04 AXIUM · Class 2 · Implant
All models of AxiumTM (139)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847536029644), received between Dec 6, 2021 and Dec 6, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Stretched” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Stretched1
- Therapy Delivered to Incorrect Body Area1
- Use of Device Problem1
Patient problems
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 6, 2021Therapy Delivered to Incorrect Body Area; Stretched; Use of Device Problem; Therapy Delivered to Incorrect Body Area; Stretched; Use of Device Problem
MEDTRONIC RECEIVED INFORMATION REPORTING THAT A COIL WAS STRETCHED AND THE WIRE WAS CUT AND LEFT IN THE PATIENT'S BODY. THAT PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A SACCULAR C6 SEGMENT ANEURYSM. THE ANEURYSM MAX DIAMETER WAS 7MM AND THE NECK DIAMETER WAS 5MM. VESSEL TORTUOSITY WAS NORMAL. IT WAS REPORTED THAT DURING THE PROCEDURE, THE QC-7-30-HELIX COIL WAS SECOND COIL TO BE USED. THERE WAS NO FRICTION OR OTHER DIFFICULTY OBSERVED DURING TESTING OR DELIVERY OF THE COIL. THE SURGEON PERFORMED SEVERAL ADJUSTMENTS SUCH AS PUSHING AND PULLING THE COIL AFTER WHICH THE COIL WAS STRETCHED BUT NOT DETACHED. THE SURGEON WITHDREW THE SPRING WIRE AND THE CATHETER TOGETHER BUT THE DISTAL END OF THE COIL REMAINED CONNECTED WITH THE COIL IMPLANTED IN THE ANEURYSM AND COULD NOT BE WITHDRAWN COMPLETELY FROM THE PATIENT'S BODY. AT THE FEMORAL ARTERY ACCESS, THE SURGEON CUT THE STRETCHED SPRING WIRE SHORE AND PART OF THE WIRE STAYED IN THE PATIENT'S BODY. IT WAS NOTED THE LENGTH OF THE SPRING WIRE IN THE BODY EXTENDED APPROXIMATELY FROM THE ANEURYSM TO THE FEMORAL ARTERY PUNCTURE SITE. SUBSEQUENTLY, THE SURGEON IMPLANTED A SOLITAIRE STENT TO PREVENT THE WIRE FROM FLOATING EXC
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 12936478 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page