Skip to content

AxiumTM QC-10-30-HELIXRecalls & reported problems

Micro Therapeutics, Inc. · COIL QC-10-30-HELIX V04 AXIUM · Class 2 · Implant

All models of AxiumTM (139)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847536029705), received between Aug 6, 2021 and Aug 6, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Separation Failure” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2021Separation Failure; Separation Failure

    MEDTRONIC RECEIVED A REPORT THAT THE THE COIL WAS PUSHED OUT OF THE SHEATH, SOAKED IN HEPARIN SALINE, AND DELIVERED INTO THE CATHETER NORMALLY. HOWEVER, THREE DETACHMENT ATTEMPTS WERE MADE WITH THE DETACHMENT DEVICE, BUT THE COIL FAILED TO DETACH. A REPLACEMENT PRODUCT WAS USED TO COMPLETE THE PROCEDURE. IT WAS INDICATED THAT ALL DEVICES WERE PREPARED AS PER THE INSTRUCTIONS FOR USE (IFU), AND NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, RUPTURED ANEURYSM OF THE C4 SEGMENT OF THE INTERNAL CAROTID ARTERY (ICA) WITH A MAX DIAMETER OF 10 MM AND A 3 MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW AND VESSEL TORTUOSITY WERE NORMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12285927 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.

  • Z-0874-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
  • Z-0873-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →

Sources and updates