AxiumTM QC-2-2-3DRecalls & reported problems
Micro Therapeutics, Inc. · COIL QC-2-2-3D V04 AXIUM · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847536029804), received between Jul 18, 2022 and Jul 18, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Separation Failure” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 18, 2022Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THE TIP OF THE ECHELON 10 MICROCATHETER CRACKED DURING THE PROCESS OF FILLING THE COIL, A COIL EXPERIENCED RESISTANCE, AND ANOTHER COIL FAILED TO DETACH. THE PATIENT WAS UNDERGOING TREATMENT FOR AN UNRUPTURED, SACCULAR ANEURYSM LOCATED IN THE C7 SEGMENT OF THE LEFT INTERNAL CAROTID ARTERY. THE MAX DIAMETER WAS 6.2MM, AND THE NECK DIAMETER WAS 5.1MM. THE PATIENT'S BLOOD FLOW WAS NORMAL, AND THEIR VESSEL TORTUOSITY WAS MODERATE. THE ACCESS VESSEL WAS THE FEMORAL ARTERY, WHICH WAS 8.1MM IN DIAMETER. IT WAS REPORTED THAT THE ECHELON COIL RUPTURED DISTALLY BUT WAS INTACT. THERE WAS WAS NO FRICTION/DIFFICULTY DURING DELIVERY, AND THERE WAS NO OTHER DAMAGE TO THE CATHETER. THE INJECTION RATE WAS SAID TO BE 100PSI. THERE WAS RESISTANCE WHEN THE SPRING COIL ENTERED THE DISTAL END OF THE ECHELON MICROCATHETER, AND IT COULD NOT BE PUSHED. THERE WAS NO DAMAGE TO THE COIL OR CATHETER. ANOTHER COIL IN THE PROCEDURE COULD NOT BE DETACHED. TWO ATTEMPTS EACH WERE MADE WITH THE INSTANT DETACHER AND THE MANUAL METHOD. A SECOND INSTANT DETACHER WAS NOT USED. THERE WERE NO ISSUES WITH THE INSTANT DETACHER. THE DEVICES WERE REPLACED, AND THE PATIENT DID NOT EXPERIENCE AN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15048410 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page