AxiumTM QC-4-12-3D (UDI 00847536029972)Recalls & reported problems
Micro Therapeutics, Inc. · COIL QC-4-12-3D V04 AXIUM · Class 2 · Implant
All models of AxiumTM (139)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00847536029972), received between Mar 2, 2020 and Dec 19, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 19, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
ADDITIONAL INFORMATION WAS RECEIVED INDICATING THERE WERE NO DETACHMENT ATTEMPTS MADE WITH THE INSTANT DETACHER OR MANUAL METHOD. THE COIL WAS NOT IMPLANTED AND HAD BEEN REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13040634 ↗MalfunctionJun 29, 2021Stretched; Stretched
MEDTRONIC RECEIVED A REPORT THAT THE GUIDE CATHETER AND MICROCATHETER WERE IN PLACE DURING THE SURGERY, BUT THE COIL WAS STRETCHED. THERE WAS NO FRICTION OR DIFFICULTY DURING DELIVERY OR TESTED, A CONTINUOUS FLUSH HAD BEEN ADMINISTERED, AND THE COIL WAS NOT REPOSITIONED DURING THE PROCEDURE. A REPLACEMENT COIL WAS USED TO COMPLETE THE PROCEDURE. IT WAS INDICATED THAT ALL DEVICES WERE PREPARED AS PER THE INSTRUCTIONS FOR USE (IFU), AND NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED ANEURYSM OF THE LEFT MIDDLE CEREBRAL ARTERY (MCA) WITH A MAX DIAMETER OF 4.5 MM AND A 3 MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MINIMAL. ADDITIONAL INFORMATION RECEIVED INDICATED THE COIL STRETCH WAS NOTICED DURING DELIVERY, AND A CAUSE OF THE STRETCH COULD NOT BE DETERMINED. AN ECHELON-10 MICROCATHETER WAS USED (LOT #: B127350).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12086303 ↗MalfunctionMar 2, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
MEDTRONIC RECEIVED A REPORT THAT THE AXIUM COIL PREMATURELY DETACHED IN THE MICROCATHETER. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, RUPTURED ANEURYSM IN THE LEFT INTERNAL CAROTID ARTERY OPHTHALMIC SEGMENT WITH A MAX DIAMETER OF 5.2MM AND A 2.1MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW AND VESSEL TORTUOSITY WERE NORMAL. IT WAS REPORTED THAT DURING DELIVERY OF THE AXIUM COIL, IT AUTOMATICALLY RELEASED IN THE MICROCATHETER IN ADVANCE, CAUSING THE COIL TO BE LEFT IN THE MICROCATHETER. THE SURGEON WITHDREW THE MICROCATHETER, THEN TOOK THE COIL OUT. THE AXIUM COIL WAS REPLACED WITH A NEW COIL TO COMPLETE THE PROCEDURE. IT WAS NOTED THAT NO SURGICAL/MEDICAL INTERVENTION WAS REQUIRED, THERE WAS NO FRICTION DURING DELIVER, THE PUSHWIRE WAS NOT BENT OR BROKEN, THE COIL WAS NOT REPOSITIONED OR ROTATED, AND A CONTINUOUS FLUSH HAD BEEN ADMINISTERED. IT WAS INDICATED THAT ALL DEVICES WERE PREPARED AS PER THE INSTRUCTIONS FOR USE (IFU), AND NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9777243 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page