AxiumTM QC-6-20-3DRecalls & reported problems
Micro Therapeutics, Inc. · COIL QC-6-20-3D V04 AXIUM · Class 2 · Implant
All models of AxiumTM (139)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00847536030039), received between Jan 12, 2022 and Jan 24, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Stretched2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 24, 2023Stretched; Stretched
MEDTRONIC RECEIVED INFORMATION REGARDING AN AXIUM COIL THAT WAS STRETCHED. THAT PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A RUPTURED SACCULAR RIGHT PARA CLINOID SEGMENT ANEURYSM. THE ANEURYSM MAX DIAMETER WAS 6MM AND THE NECK DIAMETER WAS 5MM. BLOOD FLOW WAS NORMAL. VESSEL TORTUOSITY WAS MODERATE. IT WAS REPORTED THAT THE AXIUM COIL AND ALL ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THE COIL WAS ADVANCED IN THE MICROCATHETER BUT, DURING DEPLOYMENT, DID NOT TAKE THE EXPECTED DESIRED SHAPE. THE COIL WAS WITHDRAWN AND FOUND TO BE STRETCHED. IT WAS NOTED THAT THERE WAS NO FRICTION OR DIFFICULTY DURING TESTING OR DELIVERY OF THE COIL. THE PHYSICIAN HAD TRIED TO REPOSITION THE COIL TWICE. CONTINUOUS CATHETER FLUSH WAS ADMINISTERED DURING THE PROCEDURE. THE COIL WAS REPLACED WITH ANOTHER MANUFACTURER'S PRODUCT TO COMPLETE THE PROCEDURE. THERE WAS NO HARM OR INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16237223 ↗MalfunctionOct 13, 2022Break; Stretched; Detachment of Device or Device Component; Break; Stretched; Detachment of Device or Device Component
ADDITIONAL INFORMATION WAS RECEIVED THAT THE SPRING COIL WAS PARTIALLY BROKE AND STRETCHED. THE COIL DID NOT PREMATURELY SEPARATE FROM THE DELIVERY WIRE. PART OF THE COIL WAS SEPARATED FROM THE REST. THERE WERE NO OTHER ISSUES. THERE WERE NO ISSUES THAT RESULTED IN THE DAMAGE THAT WAS FOUND. THE ENTIRE COIL WAS REMOVED FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15597918 ↗MalfunctionJan 12, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
MEDTRONIC RECEIVED INFORMATION REGARDING AN AXIUM COIL THAT DETACHED PREMATURELY. THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A RUPTURED SACCULAR RIGHT POSTERIOR COMMUNICATING ARTERY ANEURYSM. THE ANEURYSM MAX DIAMETER WAS 5.9MM AND THE NECK DIAMETER WAS 4MM. BLOOD FLOW AND VESSEL TORTUOSITY WERE NORMAL. IT WAS REPORTED THAT THE AXIUM COIL AND ALL ACCESSORY DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THE AXIUM COIL (MODEL: QC-6-20-3D, LOT: B115267) WAS SELECTED TO FORM THE HOOP. THE COIL WAS ADJUSTED TWICE DURING THE PROCESS OF FORMING THE HOOP AND DURING AN ADJUSTMENT IT WAS OBSERVED THAT THE COIL HAD DETACHED FROM THE PUSHER. THERE WAS NO FRICTION OR OTHER DIFFICULTY DURING THE DELIVERY PROCESS. THE PHYSICIAN DID NOT ROTATE THE PUSHER DURING THE PROCEDURE AND THERE WAS NO DAMAGE NOTED TO THE PUSHER. THERE WAS NO ATTEMPT MADE TO DETACH THE COIL. CONTINUOUS FLUSH WAS ADMINISTERED DURING THE PROCEDURE. THE COIL DETACHED IN THE MICROCATHETER SO THE DEVICES WERE REMOVED TOGETHER AND THE COIL WAS REPLACED WITH ANOTHER OF THE SAME MODEL AND THE OPERATION PROCEEDED WITHOUT ISSUE. THERE WAS NO HARM OR INJURY TO THE PATIENT. ANCILLARY DEVICE:
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13237151 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page