AxiumTM QC-10-30-3DRecalls & reported problems
Micro Therapeutics, Inc. · COIL QC-10-30-3D V04 AXIUM · Class 2 · Implant
All models of AxiumTM (139)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536030138), received between Apr 1, 2025 and Apr 3, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Packaging Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 27, 2022Class IIOpen, Classified
Due to incorrect size and configuration labeling of the detachable coil system.
Z-1261-2022 · Root cause: Labeling Change Control · Recall details · FDA record ↗
Most reported problems
Device problems
- Packaging Problem1
- Physical Resistance/Sticking1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 3, 2025Packaging Problem; Packaging Problem
MEDTRONIC RECEIVED A REPORT THAT THE COIL WAS PUSHED OUT AND THE DIAMETER OF THE COIL WAS FOUND TO BE LESS THAN 10MM. THE COIL WAS REMOVED AND IT WAS FOUND THAT IT WAS A 2D COIL PACKAGED AS A 3D COIL. A NEW COIL WAS PLACED AND THE ISSUE WAS RESOLVED. THE DEVICE WAS PREPARED AS INDICATED IN THE IFU. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THE EVENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED ANEURYSM WITH A MAX DIAMETER OF 10MM AND A 8MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS NORMAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21767143 ↗MalfunctionApr 1, 2025Stretched; Physical Resistance/Sticking; Stretched; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE AXIUM COIL WAS STRETCHED DURING THE DELIVERY IN THE MICROCATHETER AND COULD NOT BE PUSHED OUT SMOOTHLY. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED ANEURYSM IN THE RIGHT C6 WITH A MAX DIAMETER OF 12.6MM AND A 5.48MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MODERATE. IT WAS REPORTED THAT THE AXIUM COIL WAS STRETCHED DURING DELIVERY IN THE MICROCATHETER AND COULD NOT BE PUSHED OUT SMOOTHLY. THE DEVICES WERE PREPARED AS INDICATED IN THE IFU. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THE EVENT. ADDITIONAL INFORMATION RECEIVED REPORTED THERE WERE NO OTHER OTHER ISSUES DURING THAT PROCEDURE THAT WOULD HAVE CAUSED THE DAMAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21746183 ↗
FDA clearance and classification
Cleared through 510(k) K081465 on Aug 19, 2008 (decision: Substantially Equivalent). Submitted by Micro Therapeutics, Inc. as “Axium Detachable Coil System”. The FDA review took 84 days. FDA record ↗
The FDA regulates product code KRD (device, vascular, for promoting embolization) under 21 CFR 870.3300 as a Class II device. 21 CFR 870.3300 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
Common questions about the AxiumTM QC-10-30-3D
Has the AxiumTM QC-10-30-3D been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1261-2022 (Class II), began on Apr 27, 2022: Due to incorrect size and configuration labeling of the detachable coil system. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the AxiumTM QC-10-30-3D?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Apr 1, 2025 and Apr 3, 2025. In those reports, the most common specific device problems are “Packaging Problem” (1), “Physical Resistance/Sticking” (1) and “Stretched” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the AxiumTM QC-10-30-3D?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the AxiumTM QC-10-30-3D FDA approved?
It was cleared through 510(k) K081465 on Aug 19, 2008. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the AxiumTM QC-10-30-3D MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the AxiumTM QC-10-30-3D, and is it still sold?
The FDA's device database lists Micro Therapeutics, Inc. as the company responsible for it (brand AxiumTM, model QC-10-30-3D). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization117 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page