Axium PrimeTM APB-1.5-3-3D-ESRecalls & reported problems
Micro Therapeutics, Inc. · COIL APB-1.5-3-3D-ES V04 · Class 2 · Implant
All models of Axium PrimeTM (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536031852), received between Feb 2, 2021 and Jul 22, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Separation Failure2
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 22, 2021Separation Failure; Material Deformation; Separation Failure; Material Deformation
MEDTRONIC RECEIVED A REPORT THAT AFTER IMPLANTING THE 1.5-3ES SPRING COIL, THE COIL COULD NOT BE DETACHED AFTER 5 ATTEMPTS WITH THE INSTANT DETACHER. A SECOND INSTANT DETACHED WAS NOT USED, AND THE MANUAL DETACHMENT METHOD WAS PERFORMED TWICE, BUT DETACHMENT WAS STILL UNSUCCESSFUL. IT WAS STATED THERE WAS NO ISSUE WITH THE INSTANT DETACHER. A REPLACEMENT COIL (1.5-2ES) WAS USED AND DETACHED SMOOTHLY. IT WAS INDICATED THAT ALL DEVICES WERE PREPARED AS PER THE INSTRUCTIONS FOR USE (IFU), AND NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A FUSIFORM, UNRUPTURED ANEURYSM OF THE OPHTHALMIC ARTERY SEGMENT WITH A MAX DIAMETER OF 4.2 MM AND A 3.7 MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MODERATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12211315 ↗MalfunctionFeb 2, 2021Separation Failure; Separation Failure
ADDITIONAL INFORMATION RECEIVED THAT THREE DETACHMENTS HAD BEEN MADE WITH THE INSTANT DETACHER, AND ONE MANUAL DETACHMENT ATTEMPT HAD BEEN MADE WITH A PULL OF THE INNER WIRE. PRIOR TO THE NON-DETACHMENT THERE HAD NOT BEEN ANY ISSUES, AND NO DAMAGE WAS OBSERVED TO THE PUSHWIRE AFTER REMOVAL. THE DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE, WITH THE ONLY DEVIATION THAT A HEADWAY DUO 167CM .013 ID CATHETER WAS USED. THE PATIENT WAS BEING TREATED FOR FUSIFORM VESSEL SACRIFICE IN THE POSTERIOR INFERIOR CEREBELLAR ARTERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11267241 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page