Axium PrimeTM APB-2.5-6-3D-ESRecalls & reported problems
Micro Therapeutics, Inc. · COIL APB-2.5-6-3D-ES V04 · Class 2 · Implant
All models of Axium PrimeTM (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00847536031913), received between Nov 18, 2020 and Nov 1, 2021 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Separation Failure5
- Physical Resistance/Sticking2
- Break1
- Detachment of Device or Device Component1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 1, 2021Separation Failure; Separation Failure
ADDITIONAL INFORMATION RECEIVED REPORTED THAT PRIOR TO COIL NON-DETACHMENT THERE WERE NO ISSUE ENCOUNTERED. THERE WAS NO PUSHWIRE DA MAGE OBSERVED AFTER REMOVAL FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12728568 ↗MalfunctionJun 8, 2021Separation Failure; Separation Failure
MEDTRONIC RECEIVED INFORMATION AN AXIUM COIL THAT FAILED TO DETACH DURING A PROCEDURE. THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT AN UNRUPTURED SACCULAR ANEURYSM OF A M2 SEGMENT. PATIENT VESSEL TORTUOSITY WAS VERY SEVERE AND THE PHYSICIAN REPORTED IT MADE THE CASE DIFFICULT. IT WAS REPORTED THAT ALL DEVICES WERE PREPARED AS INDICATED IN THE INSTRUCTIONS FOR USE (IFU). THE CATHETER WAS FLUSHED CONTINUOUSLY DURING THE PROCEDURE. THE MICROCATHETER WAS PLACED AT THE INTENDED LOCATION IN THE M2. THE AXIUM COIL (MODEL: APB-2.5-6-3D-ES, LOT: B093781) WAS DELIVERED TO THE ANEURYSM BUT COULD NOT BE DETACHED. ABOUT 10 ATTEMPTS WERE MADE TO DETACH THE COIL WITH THE INSTANT DETACHER (ID) WITHOUT SUCCESS. A SECOND ID WAS TRIED, AGAIN ABOUT 10 TIMES, WITHOUT SUCCESS. IT WAS NOTED THERE WAS NOTHING WRONG WITH EITHER ID. MANUAL DETACHMENT WAS ALSO ATTEMPTED 5 TIMES WITHOUT SUCCESS. FINALLY THE COIL WAS REMOVED AND REPLACED. THERE WAS NO DAMAGE NOTED TO THE CATHETER. THIS COIL REPLACED. THERE WAS NO HARM OR INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11960106 ↗MalfunctionJun 7, 2021Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE COIL PUSHED OUT OF THE INTRODUCER SHEATH SMOOTHLY, AND THERE WERE NO ISSUES THAT RESULTED IN THE DAMAGE THAT WAS FOUND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11950255 ↗MalfunctionJun 2, 2021Separation Failure; Separation Failure
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THERE WERE NO ISSUES ENCOUNTERED PRIOR TO THE NON-DETACHMENT, AND NO DAMAGE WAS OBSERVED TO THE PUSHWIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11925342 ↗MalfunctionDec 8, 2020Separation Failure; Separation Failure
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE PHYSICIAN ATTEMPTED TO DETACH 2 OR 3 TIMES USING AN INSTANT DETACHER AND THE MANUAL DETACHMENT METHOD, BUT THOUGHT IT WOULDN'T WORK BECAUSE THE COIL WAS BROKEN. PRIOR TO NON-DETACHMENT THE COIL WAS BROKEN IN THE MIDDLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10969154 ↗MalfunctionDec 7, 2020Detachment of Device or Device Component; Material Deformation; Physical Resistance/Sticking; Detachment of Device or Device Component; Material Deformation; Physical Resistance/Sticking
NO INVESTIGATION REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10959193 ↗MalfunctionNov 18, 2020Separation Failure; Separation Failure
NO ADDITIONAL INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10864477 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page