Axium PrimeTM APB-4-6-3D-SSRecalls & reported problems
Micro Therapeutics, Inc. · COIL APB-4-6-3D-SS V04 · Class 2 · Implant
All models of Axium PrimeTM (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00847536031982), received between Oct 7, 2020 and Dec 7, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Break1
- Physical Resistance/Sticking1
- Separation Failure1
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 7, 2022Break; Stretched; Detachment of Device or Device Component; Break; Stretched; Detachment of Device or Device Component
MEDTRONIC RECEIVED A REPORT REGARDING PHENOM CATHETER RESISTANCE AND AXIUM COIL PREMATURE DETACHMENT. THE PATIENT WAS UNDERGOING INTRACRANIAL ANEURYSM TREATMENT OF A SACCULAR, UNRUPTURED ANEURYSM IN THE RIGHT SIDE OF THE OPHTHALMIC ARTERY WITH A MAX DIAMETER OF 6 MM AND A 5 MM NECK DIAMETER. THE ACCESSED VESSEL WAS THE FEMORAL ARTERY WITH A DIAMETER OF 9 MM. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS NORMAL. IT WAS REPORTED THAT THE ANEURYSM EMBOLIZATION SURGERY WAS PERFORMED AND THE SHAPEABLE MICROCATHETER PHENOM PLUS WAS SELECTED TO ESTABLISH THE PATHWAY. THE MICROCATHETER COULD NOT BE PUSHED OUT. IT WAS REPORTED THE CATHETER RESISTANCE OCCURRED IN THE DISTAL END. IT WAS ALSO REPORTED THE AXIUM COIL WAS SELECTED AND USED TO FORM A HOOP. THE SPRING COIL WAS STRETCHED. IT IS NOTED THAT COIL SEPARATION/BREAK/PREMATURE DETACHMENT OCCURRED. THE COIL WAS REMOVED FROM THE PATIENT. THE COIL WAS STRETCH AND WAS IMMEDIATELY TAKEN OUT. NO SURGICAL OR MEDICINAL INTERVENTION WAS REQUIRED. THE PUSHWIRE WAS NOT BENT OR BROKEN. THERE WAS NO FRICTION OR DIFFICULTY DURING DELIVERY OBSERVED. THE PHYSICIAN DID NOT REPOSITION THE COIL. THE PHYSICIAN DID NOT ATTEMPT TO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15936168 ↗MalfunctionNov 7, 2021Separation Failure; Separation Failure
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THERE WERE NO ISSUES ENCOUNTERED PRIOR TO THE NON-DETACHMENT. THERE WERE TWO MANUAL DETACHMENTS THAT WERE UNSUCCESSFUL AND IT WAS DETACHED WHEN IT WAS WITHDRAWN AS A WHOLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12759883 ↗MalfunctionOct 7, 2020Detachment of Device or Device Component; Physical Resistance/Sticking; Detachment of Device or Device Component; Physical Resistance/Sticking
IT WAS REPORTED THAT THE FIRST AXIUM COIL EXPERIENCED RESISTANCE DURING DELIVERY IN THE ECHELON MICROCATHETER. WHEN THE PHYSICIAN PULLED BACK ON THE COIL, THE COIL WAS DETACHED INSIDE THE CATHETER. IT WAS NOTED REPOSITIONING WAS PERFORMED 2-3 TIMES, NO DETACHMENT ATTEMPTS WERE MADE, AND A CONTINUOUS FLUSH HAD BEEN ADMINISTERED. SINCE THE COIL WAS INSIDE THE CATHETER, BOTH DEVICES WERE REMOVED TOGETHER. A SECOND COIL AXIUM COIL WAS THEN DELIVERED, BUT WHEN THE PHYSICIAN TRIED TO PULLTHE COIL BACK OUT OF THE VESSEL, IT COMPLETELY DETACHED: PART IN THE A1 AND PART IN THE M1. THE COIL WAS FLUSHED IN THE M3 SEGMENT, AND A SOLITAIRE RETRIEVER WAS USED TO REMOVE THE COIL FROM THE PATIENT. IT WAS NOTED REPOSITIONING WITH THIS COIL WAS DONE 2-3 TIMES, ROTATION OF THE PUSHWIRE WAS DONE, AND A CONTINUOUS FLUSH WAS ALSO ADMINISTERED. REPLACEMENT DEVICES WERE USED TO COMPLETE THE PROCEDURE. IT WAS INDICATED THAT ALL DEVICES WERE PREPARED AS PER THE INSTRUCTIONS FOR USE (IFU), THE PATIENT'S BLOOD FLOW AND VESSEL TORTUOSITY WERE NORMAL, AND NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10640098 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.