Axium PrimeTM APB-3-8-HX-ESRecalls & reported problems
Micro Therapeutics, Inc. · COIL APB-3-8-HX-ES V04 · Class 2 · Implant
All models of Axium PrimeTM (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00847536032279), received between Nov 22, 2019 and Dec 1, 2021 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Separation Failure4
- Detachment of Device or Device Component1
- Material Deformation1
- Physical Resistance/Sticking1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 1, 2021Detachment of Device or Device Component; Material Deformation; Physical Resistance/Sticking; Detachment of Device or Device Component; Material Deformation; Physical Resistance/Sticking
ADDITIONAL INFORMATION RECEIVED REPORTED THE COIL PUSHED OUT OF THE INTRODUCER SHEATH SMOOTHLY. NO ATTEMPT WAS MADE AT DETACHMENT; THE COIL WAS NEVER PUSHED OUT OF THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12909015 ↗MalfunctionMay 13, 2021Separation Failure; Separation Failure
ADDITIONAL INFORMATION RECEIVED INDICATED COIL DETACHMENT WAS ATTEMPTED THREE TIMES WITH THE INSTANT DETACHER (ID), AND THE MANUAL DETACHMENT METHOD WAS PERFORMED ONCE. THERE WERE NO OTHER ISSUES ENCOUNTERED PRIOR TO THE NON-DETACHMENT, AND THERE WAS NO DAMAGE OBSERVED TO THE PUSHWIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11816408 ↗MalfunctionNov 22, 2019Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THREE AXIUM COILS COULD NOT BE DETACHED USING 2 DIFFERENT INSTANT DETACHERS. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN ANEURYSM. IT WAS REPORTED THAT THE COILS COULD NOT BE DETACHED IN ANEURYSM. THE USE OF A NEW DETACHER DID NOT WORK. THE COIL WAS WITHDRAWN FROM THE ANEURYSM, A SECOND COIL WAS USED INSTEAD WHICH ALSO COULD NOT BE DETACHED. THE SECOND COIL WAS ALSO WITHDRAWN AND REPLACED BY A THIRD COIL WHICH COULD ALSO NOT BE DETACHED. IT WAS UNKNOWN HOW MANY ATTEMPTS WERE PERFORMED TO DETACHED THE COIL WITH THE INSTANT DETACHER. A MANUAL ATTEMPT WAS NOT DONE VIA THE HYPOTUBE. IT WAS NOTED ALL PRODUCTS WERE PREPARED AS PER THE INSTRUCTIONS FOR USE (IFU), AND NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. REFER TO MANUFACTURER REPORT 2029214-2019-01157 FOR DETAILS PERTAINING TO THE RELATED REPORTABLE EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9362376 ↗MalfunctionNov 22, 2019Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THREE AXIUM COILS COULD NOT BE DETACHED USING 2 DIFFERENT INSTANT DETACHERS. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN ANEURYSM. IT WAS REPORTED THAT THE COILS COULD NOT BE DETACHED IN ANEURYSM. THE USE OF A NEW DETACHER DID NOT WORK. THE COIL WAS WITHDRAWN FROM THE ANEURYSM, A SECOND COIL WAS USED INSTEAD WHICH ALSO COULD NOT BE DETACHED. THE SECOND COIL WAS ALSO WITHDRAWN AND REPLACED BY A THIRD COIL WHICH COULD ALSO NOT BE DETACHED. IT WAS UNKNOWN HOW MANY ATTEMPTS WERE PERFORMED TO DETACHED THE COIL WITH THE INSTANT DETACHER. A MANUAL ATTEMPT WAS NOT DONE VIA THE HYPOTUBE. IT WAS NOTED ALL PRODUCTS WERE PREPARED AS PER THE INSTRUCTIONS FOR USE (IFU), AND NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. REFER TO MANUFACTURER REPORT 2029214-2019-01157 FOR DETAILS PERTAINING TO THE RELATED REPORTABLE EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9362326 ↗MalfunctionNov 22, 2019Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THREE AXIUM COILS COULD NOT BE DETACHED USING 2 DIFFERENT INSTANT DETACHERS. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF AN ANEURYSM. IT WAS REPORTED THAT THE COILS COULD NOT BE DETACHED IN ANEURYSM. THE USE OF A NEW DETACHER DID NOT WORK. THE COIL WAS WITHDRAWN FROM THE ANEURYSM, A SECOND COIL WAS USED INSTEAD WHICH ALSO COULD NOT BE DETACHED. THE SECOND COIL WAS ALSO WITHDRAWN AND REPLACED BY A THIRD COIL WHICH COULD ALSO NOT BE DETACHED. IT WAS UNKNOWN HOW MANY ATTEMPTS WERE PERFORMED TO DETACHED THE COIL WITH THE INSTANT DETACHER. A MANUAL ATTEMPT WAS NOT DONE VIA THE HYPOTUBE. IT WAS NOTED ALL PRODUCTS WERE PREPARED AS PER THE INSTRUCTIONS FOR USE (IFU), AND NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9362304 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page