Axium PrimeTM APB-4-12-HX-SSRecalls & reported problems
Micro Therapeutics, Inc. · COIL APB-4-12-HX-SS V04 · Class 2 · Implant
All models of Axium PrimeTM (66)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00847536032323), received between Sep 2, 2020 and May 24, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Migration” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Migration1
- Physical Resistance/Sticking1
- Stretched1
Patient problems
- Device Embedded In Tissue or Plaque1
- Dysphasia1
- Encephalopathy1
- Hemorrhage, Subarachnoid1
- Muscle Weakness/Atrophy1
- No Clinical Signs, Symptoms or Conditions1
- Vascular Dissection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 24, 2022Physical Resistance/Sticking; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THE COIL WAS STUCK IN THE PROXIMAL SECTION OF THE CATHETER. THE IMPLANT COIL PUSHED SMOOTHLY OUT FROM THE INTRODUCER SHEATH. THERE WAS NO COIL DAMAGE. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. ANCILLARY DEVICES INCLUDE A NAVIEN072 GUIDE CATHETER, SL-10 MICROCATHETER, AND SYNCHRO14 GUIDEWIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14489928 ↗InjurySep 2, 2020Stretched; Migration; Stretched; Migration
MEDTRONIC RECEIVED A REPORT THAT THE AXIUM COIL MIGRATED, BECAME STUCK ON THE PIPELINE AND STRETCHED RESULTING IN THE COIL BEING LEFT IN THE PATIENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A RECURRENT SACCULAR, RUPTURED ANEURYSM OF THE LEFT, POSTERIOR COMMUNICATING (PCOM) ARTERY WITH A MAX DIAMETER OF 8MM AND A 4MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW AND VESSEL TORTUOSITY WERE NORMAL. IT WAS REPORTED THAT THE PLAN IN THE CASE WAS TO LAY A SECOND LAYER OF PIPELINE IN THE INTERNAL CAROTID ARTERY AND THEN APPROACH THE ANEURYSM FROM THE POSTERIOR FLOW AND COIL THE RECAN AND COIL OCCLUDE THE POSTERIOR COMMUNICATING (PCOM) ARTERY. THE PIPELINE DEPLOYMENT WENT WITHOUT ISSUE. THE MICROCATHETER WAS ABLE TO BE DELIVERED AND PLACED JUST OUTSIDE OF THE ANEURYSM. THE AXIUM SOFT COIL WAS DELIVERED, AND IT WAS THOUGHT THE COIL SHOULD HAVE BEEN PROTECTED FROM ENTERING THE ICA BY THE DOUBLE PIPELINES. THE AXIUM COIL SEEMED TO FILL THE ANEURYSM, BUT THEN MIGRATED PAST THE PIPELINE AND INTO THE MIDDLE CEREBRAL ARTERY (MCA). IT WAS NOTED NO REAL FORCE WAS REQUIRED. DYNA CT APPEARED TO SHOW THE COIL WENT THROUGH THE PIPELINE, AND WAS FLOATING WITHIN THE STENT AND OUT INTO
Patient problems: Device Embedded In Tissue or Plaque; Dysphasia; Encephalopathy; Hemorrhage, Subarachnoid; Muscle Weakness/Atrophy; Vascular Dissection
MAUDE report 10486439 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.