Skip to content

AxiumTM PC-2-2-HELIXRecalls & reported problems

Micro Therapeutics, Inc. · COIL PC-2-2-HELIX V04 · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847536032941), received between Apr 30, 2021 and Apr 30, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Separation Failure” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 30, 2021Separation Failure; Separation Failure

    IT WAS REPORTED THE COIL COULD NOT BE DETACHED IN THE PATIENT'S BODY. THE PHYSICIAN ATTEMPTED 3 TIMES TO DETACH THE COIL WITH THE INSTANT DETACHER. ANOTHER INSTANT DETACHER WAS USED AND AGAIN 3 ATTEMPTS WERE MADE TO DETACH THE COIL. 1 MANUAL ATTEMPT VIA HYPOTUBE TO DETACH THE COIL WAS MADE. IT WAS NOTED THAT THERE WERE NO ISSUES WITH THE INSTANT DETACHER. IT WAS NOTED THE VESSEL TORTUOSITY WAS MODERATE. THERE WERE NO PATIENT SYMPTOMS ASSOCIATED WITH THE EVENT. THE DEVICES WERE PREPARED AS INDICATED IN THE IFU. IT WAS NOTED THAT THE PATIENT INJURY DETAILS SAID ANEURYSM; HOWEVER, COILS DO NOT CAUSE ANEURYSMS. ANXILLARY DEVICE: ECHELON 10 MICROCATHETER

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11753826 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.

  • Z-0873-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
  • Z-0874-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page