AxiumTM PC-4-8-HELIXRecalls & reported problems
Micro Therapeutics, Inc. · COIL PC-4-8-HELIX V04 · Class 2 · Implant
All models of AxiumTM (139)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847536033023), received between Mar 30, 2020 and Mar 30, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Physical Resistance/Sticking1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2020Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THREE AXIUM COILS EXPERIENCED RESISTANCE DURING DELIVERY, AND ONE BROKE UPON RETRIEVAL. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A CCF NEUROVASCULAR ABNORMALITY. IT WAS NOTED THE PATIENT'S BLOOD FLOW AND VESSEL TORTUOSITY WERE NORMAL. IT WAS REPORTED THAT THE REBAR MICROCATHETER WAS PLACED, AND THE FIRST COIL WAS DEPLOYED WITH NO ISSUE. WHEN DELIVERING THE SECOND COIL (PLI10), THERE WAS A STRONG RESISTANCE FELT, BUT IT WAS DEPLOYED SUCCESSFULLY. THE THIRD COIL (PLI20) EXPERIENCED STRONGER RESISTANCE AND COULD NOT BE DEPLOYED. DURING RETRIEVAL THE COIL BROKE INTO TWO PIECES; ONE PIECE WAS INSIDE THE ANEURYSM AT THE INTENDED LOCATION, THE OTHER PIECE WAS RETRIEVED SUCCESSFULLY, AND THE PUSHWIRE WAS BENT/BROKEN UPON INSPECTION. THE REPLACEMENT COIL DEPLOYED EASILY AND IN TOTAL, SIX COILS WERE USED TO FILL THE ANEURYSM. A SEVENTH COIL (PLI30) WAS TO BE ADDED, BUT RESISTANCE WAS FELT AGAIN DURING THE DELIVERY. THE FLOW WAS CHECKED WITHOUT THE COIL, AND IT WAS DETERMINED THE LAST COIL WAS NOT NEEDED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT HARM INDICATED IN THE EVENT. ANCILLARY DEVICES INCLUDE A REBAR 18 MICROCATHETER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9898301 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page