AxiumTM PC-9-30-HELIXRecalls & reported problems

Micro Therapeutics, Inc. · COIL PC-9-30-HELIX V04 · Class 2 · Implant

All models of AxiumTM (139)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847536033139), received between Apr 27, 2022 and Apr 27, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Separation Failure” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 27, 2022Separation Failure; Separation Failure

    MEDTRONIC RECEIVED INFORMATION THAT AN AXIUM COIL FAILED TO DETACH. THE PATIENT WAS BEING TREATED FOR NEUROVASCULAR ABNORMALITIES THAT WERE NOT IN A ELOQUENT REGION. VESSEL TORTUOSITY WAS NORMAL. THE DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU).   THE COILS CANNOT DETACHED EVEN CHANGE NEW DETACHER SO CUSTOMER DECIDED TO REMOVE THE COIL AND REPLACE WITH NEW COIL INSTEAD. THE PHYSICIAN ATTEMPTED TO DETACH WITH THE INSTANT DETACHER 3 TIMES. THEY TRIED ANOTHER DETACHER TWO TIMES. NO MANUAL ATTEMPT TO DETACH WAS MADE. THERE WAS NO ISSUE WITH THE INSTANT DETACHER. NO SYMPTOMS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14221938 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.

  • Z-0874-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
  • Z-0873-2012 · Mar 1, 2010 · Terminated
    Axium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab

Other Device, Vascular, for Promoting Embolization devices

All Device, Vascular, for Promoting Embolization devices →