ConcertoTM NV-4-8-HELIX (UDI 00847536033535)Recalls & reported problems
Micro Therapeutics, Inc. · COIL NV-4-8-HELIX V06 · Class 2 · Implant
All models of ConcertoTM (70)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00847536033535), received between Oct 21, 2020 and Jul 27, 2022 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (5 reports).
2 other recalls cover devices cleared under the same FDA submission (K081465); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Separation Failure5
- Adverse Event Without Identified Device or Use Problem1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 27, 2022Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THE CONCERTO COIL DID NOT DETACH. THE COIL WAS REPLACED, AND THE PATIENT DID NOT EXPERIENCE ANY INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15118088 ↗MalfunctionApr 27, 2022Separation Failure; Separation Failure
MEDTRONIC RECEIVED INFORMATION THAT A CONCERTO COIL WOULD NOT DETACH. THE PATIENT WAS BEING TREATED FOR A GASTRO INTESTINAL BLEED. PATIENT BLOOD FLOW AND VESSEL TORTUOSITY WERE NORMAL. THE DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). COIL WOULD NOT DETACH WHEN OUTSIDE MICROCATHETER. THE PHYSICIAN ATTEMPTED TO DETACH WITH THE INSTANT DETACHER 2 TIMES. THEY DID NOT TRY ANOTHER INSTANT DETACHER OR ATTEMPT TO DETACH MANUALLY. THERE WAS NO ISSUE WITH THE INSTANT DETACHER. NO SYMPTOMS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14222242 ↗InjuryOct 6, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED INFORMATION THAT THERE WERE TWO CASES OF COILS THAT ACHIEVED HEMOSTASIS AND THE PATIENTS CAME BACK ABOUT A WEEK LATER WITH FLOW THROUGH THE COIL PACK. ADDITIONAL INFORMATION RECEIVED REPORTED 3 CONCERTO COILS WERE INVOLVED IN THE EVENT. THE INITIAL EMBOLIZATION PROCEDURE WAS COMPLETED WITH (B)(6) 2021; FLOW THROUGH THE COILS WAS OBSERVED AND THE PATIENT WAS SYMPTOMATIC. A SECOND EMBOLIZATION PROCEDURE OF THE SAME ARTERY WAS COMPLETED (B)(6) 2021. NO IMAGES WERE AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 12586467 ↗MalfunctionJul 20, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT THE END OF THE COIL WIRE WAS BENT. ADDITIONAL INFORMATION RECEIVED REPORTED THERE WAS DAMAGE NOTICED UPON OPENING THE PACKAGE, AND PREPARATION WAS PERFORMED PRIOR TO THE DAMAGE BEING SEEN. THERE WERE NO ISSUES THAT RESULTED IN THE DAMAGE THAT WAS FOUND.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12200645 ↗MalfunctionMay 27, 2021Separation Failure; Separation Failure
ADDITIONAL INFORMATION RECEIVED REPORTED THERE WERE NO ISSUE DURING THE PROCEDURE PRIOR TO NON-DETACHMENT. THERE WAS NO DAMAGE OBSERVED TO THE PUSHWIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11898213 ↗MalfunctionMay 4, 2021Separation Failure; Separation Failure
ADDITIONAL INFORMATION RECEIVED FROM A HEALTHCARE PROVIDER (HCP), VIA A MANUFACTURING REPRESENTATIVE (REP), INDICATED THE FOLLOWING WERE ALL UNKNOWN: HOW MANY DETACHMENT ATTEMPTS WERE MADE VIA INSTANT DETACHER, THE NUMBER OF DETACHMENT ATTEMPTS VIA MANUAL METHOD, LOT NUMBER OF INSTANT DETACHER, IF THERE ANY ISSUES PRIOR TO NON-DETACHMENT, AND IF THERE WAS ANY DAMAGE OBSERVED TO THE DEVICE AFTER REMOVAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11769584 ↗MalfunctionOct 21, 2020Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THE PHYSICIAN ATTEMPTED TO DEPLOY THE CONCERTO COIL TWICE WITH THE INSTANT DETACHER AS WELL AS MANUALLY, BUT THE COIL WOULD NOT DETACH. THE COIL WAS RETRIEVED SAFELY FROM THE PATIENT AND REPLACED WITH ANOTHER PRODUCT OF THE SAME SIZE TO COMPLETE THE PROCEDURE. IT WAS NOTED A SECOND INSTANT DETACHER WAS NOT USED, AND THERE WAS NO ISSUE WITH THE FIRST INSTANT DETACHER. IT WAS INDICATED THAT ALL DEVICES WERE PREPARED AS PER THE INSTRUCTIONS FOR USE (IFU), AND NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF MESENTERIC ARTERY EMBOLIZATION. IT WAS NOTED THE PATIENT'S VESSEL TORTUOSITY WAS MINIMAL.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10709394 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081465. One submission often covers many models, so check the product and lot details in each record.
- Z-0874-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular a
- Z-0873-2012 · Mar 1, 2010 · TerminatedAxium Detachable Coil System QC-2-1-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular ab
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page