ConcertoTM NV-10-30-HELIX (UDI 00847536033610)Recalls & reported problems
Micro Therapeutics, Inc. · COIL NV-10-30-HELIX V06 · Class 2 · Implant
All models of ConcertoTM (70)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00847536033610), received between Jan 13, 2021 and Oct 25, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Detachment of Device or Device Component2
- Separation Failure2
- Material Deformation1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 25, 2023Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THE COIL CAN'T BE DETACHED. THE DEVICE AND ANY ACCESSORIES WERE PREPARED AS INDICATED IN THE IFU. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18006029 ↗MalfunctionSep 2, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
MEDTRONIC RECEIVED A REPORT THAT THE CONCERTO COIL DETACHED BY ITSELF FROM THE WIRE. A SNARE WAS USED TO REMOVE THE COIL, AND A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE PATIENT DID NOT EXPERIENCE ANY INJURY OR COMPLICATIONS. THE DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15351273 ↗MalfunctionApr 13, 2022Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THE CONCERTO COIL WAS UNABLE TO BE DETACHED WITH THE INSTANT DETACHER AND THE MANUAL DETACHMENT METHOD. THE DEVICE WAS PREPARED AS INDICATED IN THE IFU WITHOUT ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14092724 ↗MalfunctionDec 2, 2021Break; Detachment of Device or Device Component; Material Deformation; Physical Resistance/Sticking; Break; Detachment of Device or Device Component; Material Deformation; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THERE WAS RESISTANCE FELT DURING DELIVERY AND THE COIL WAS BROKEN. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SACCULAR, UNRUPTURED ANEURYSM IN THE SUBCARINAL VASCULAR LOCATION WITH AN UNKNOWN MAX DIAMETER AND A 3MM NECK DIAMETER. IT WAS NOTED THE PATIENT'S BLOOD FLOW WAS HIGH AND VESSEL TORTUOSITY WAS SEVERE. IT WAS REPORTED THAT THE DOCTORS FOUND A PIECE OF BROKEN COIL WIRE WHEN THEY REMOVED THE INTRODUCER SHEATH FROM THE FEMORAL ARTERY AFTER COMPLETING THE PROCEDURE. SO THEY ARRANGED THE PATIENT TO DO ANOTHER CT SCAN AND THEY FOUND THAT THE COIL WIRE/FILAMENT WAS BROKEN AND RETAINED IN THE PATIENT. IT WAS DECIDED TO CUT DOWN THE GROIN AND ATTEMPT TO USE A SNARE AND REMOVE THE COIL FILAMENT. AFTER A FEW ATTEMPTS, PART OF THE COIL FILAMENT WAS REMOVED AND BROKEN INTO THREE PIECES. THERE WAS A SHORT SEGMENT OF THE COIL FILAMENT THAT COULD NOT BE REMOVED AND REMAINED IN THE PATIENT'S BODY. THE PATIENT'S CONDITION WAS FINE AND THERE WAS NO FURTHER TREATMENT NEEDED AT THE MOMENT. IT WAS THOUGHT THAT THE COIL FILAMENT WAS MAYBE DUE TO A PREMATURE DETACHMENT OF THE COIL DURING DELIVERY. THERE WAS STRONG RESISTANCE FELT DURING DELIVERY AND THE C
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 12915959 ↗MalfunctionJan 13, 2021Break; Break
NO ADDITIONAL INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11167150 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page