- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00847536033665), received between Oct 6, 2020 and May 4, 2021 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Migration” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Migration2
- Positioning Problem2
- Separation Failure2
Patient problems
- Aneurysm2
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 4, 2021Separation Failure; Separation Failure
IT WAS REPORTED THE COIL WAS BEING USED AS PER THE IFU, BUT IT COULD NOT BE DETACHED EITHER INSTANT DETACHER NOR MANUAL METHOD. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11769939 ↗InjuryMar 5, 2021Positioning Problem; Migration; Positioning Problem; Migration
ADDITIONAL INFORMATION RECEIVED REPORTED THE PATIENT DID NOT RECEIVE ANY ANTIPLATELET REGIMEN.
Patient problems: Aneurysm
MAUDE report 11424080 ↗InjuryMar 5, 2021Positioning Problem; Migration; Positioning Problem; Migration
ADDITIONAL INFORMATION RECEIVED REPORTED THAT REVIEW OF THE CASE WITH THE VASCULAR SURGEON AND INTERVENTIONAL RADIOLOGIST FOUND THAT THERE WAS NO MIGRATION OF ANY OF THE COILS. THEY WERE PLACED AS THE PHYSICIAN POSITIONED THEM DURING THE PROCEDURE TO ADDRESS THE RUPTURED ANEURYSM. HOWEVER, THE PHYSICIAN POSITIONED THE COILING "TOO DOWNSTREAM," RESULTING IN COLLATERAL FEEDERS STILL PROVIDING FLOW INTO THE ANEURYSM SAC. THE RADIOLOGIST CLOSED THE ENDOLEAK WITH ONYX INJECTION.
Patient problems: Aneurysm
MAUDE report 11424078 ↗MalfunctionOct 6, 2020Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THE HEALTHCARE PROVIDER (HCP) WROTE "DEFECTIVE - WON'T DEPLOY" ON THE OUTSIDE OF THE AXIUM COIL BOX. IT WAS BELIEVED THE COIL WOULD NOT DETACH, BUT THE MANUFACTURING REPRESENTATIVE (REP) DID NOT HAVE A CLEAR UNDERSTANDING OF WHAT HAPPENED. THERE WAS NO INJURY TO THE PATIENT WITH THE EVENT. THE PATIENT WAS UNDERGOING SURGERY FOR TREATMENT OF A SPLENIC SAC EMBOLIZATION.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10634594 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗