ConcertoTM PV-3-8-3D (UDI 00847536033740)Recalls & reported problems
Micro Therapeutics, Inc. · COIL PV-3-8-3D V04 · Class 2 · Implant
All models of ConcertoTM (70)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00847536033740), received between May 26, 2021 and Nov 3, 2022 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Physical Resistance/Sticking4
- Separation Failure4
- Material Deformation2
- Detachment of Device or Device Component1
- Inadequacy of Device Shape and/or Size1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 3, 2022Inadequacy of Device Shape and/or Size; Separation Failure; Physical Resistance/Sticking; Inadequacy of Device Shape and/or Size; Separation Failure; Physical Resistance/Sticking
ADDITIONAL INFORMATION WAS RECEIVED THAT THE COILS GOT STUCK IN THE MICROCATHETER TO CATHARIZE THE GASTRODUODENAL ARTERY USING A 0.018 WIRE. THE COILS WERE REMOVED FROM THE MICROCATHETER AND DID NOT CAUSE HARM TO THE PATIENT FROM EPIC. WHILE ADVANCING THE COIL PRIOR TO DEPLOYMENT, THE COIL MALFUNCTIONED WITHIN THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15721545 ↗MalfunctionOct 27, 2022Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation
MEDTRONIC RECEIVED A REPORT THAT ONE CONCERTO COIL BECAME STUCK IN THE INTRODUCER SHEATH, AND ANOTHER CONCERTO COIL BECAME DAMAGED. IT WAS REPORTED THAT THE SECOND COIL USED COULD NOT BE ADVANCED OUT OF THE INTRODUCER SHEATH. ANOTHER COIL USED IN THE PROCEDURE WAS "NOT HUBBED UP" WHEN IT WAS BEING FED. THEY TRIED TO RECAPTURE THE COIL BUT IT WAS MANGLED. THE EVENT WAS SAID TO BE NOT ASSOCIATED WITH THE PATIENT. THE DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). ANCILLARY DEVICES INCLUDE A REBAR 18 MICROCATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15680913 ↗MalfunctionSep 20, 2022Separation Failure; Material Deformation; Physical Resistance/Sticking; Separation Failure; Material Deformation; Physical Resistance/Sticking
ADDITIONAL INFORMATION RECEIVED REPORTED THAT PATIENT'S VESSEL TORTUOSITY WAS MODERATE, AND THE COILS HAD COME OUT OF THE INTRODUCER SHEATH SMOOTHLY. THE MICROCATHETER USED IN THE EVENT WAS A PROGREAT 2.8FR. NO DETACHMENT ATTEMPTS WERE MADE AS THE COILS GOT STUCK INSIDE THE MICROCATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15456875 ↗MalfunctionMay 13, 2022Separation Failure; Physical Resistance/Sticking; Separation Failure; Physical Resistance/Sticking
MEDTRONIC RECEIVED A REPORT THAT THERE WAS RESISTANCE WITH 4 CONCERTO COILS AND THE COIL WOULDN'T DETACH. THE PATIENT WAS UNDERGOING TREATMENT FOR PROSTATIC ARTERY EMBOLIZATION (PAE). THE PATIENT'S BLOOD FLOW AND VESSEL TORTUOSITY WERE NORMAL. IT WAS REPORTED THAT TWO COILS WERE STUCK IN THE PROXIMAL PART OF THE MICROCATHETER, ONE WAS STUCK IN THE CATHETER HUB, AND ONE WAS STUCK IN THE DISTAL PART OF THE CATHETER. THE COILS HAD PUSHED OUT OF THE INTRODUCER SHEATH SMOOTHLY. THERE WAS NO DAMAGE TO THE CATHETER OR COILS. THE PATIENT DID NOT EXPERIENCE ANY INJURY OR COMPLICATIONS. THE DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14399492 ↗MalfunctionDec 7, 2021Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THE COIL WOULD NOT DEPLOY. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12956900 ↗MalfunctionMay 26, 2021Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THE COIL WOULD NOT MOVE INTO THE MICROCATHETER FROM THE INTRODUCER SHEATH. THE PATIENT WAS UNDERGOING SURGERY FOR A Y90 EMBOLIZATION. IT WAS NOTED THE VESSEL TORTUOSITY WAS MODERATE. THERE WERE NO PATIENT SYMPTOMS ASSOCIATED WITH THE EVENT. THE DEVICES WERE PREPARED AS INDICATED IN THE IFU.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11882332 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page