ConcertoTM PV-14-40-3D (UDI 00847536033849)Recalls & reported problems
Micro Therapeutics, Inc. · COIL PV-14-40-3D V04 · Class 2 · Implant
All models of ConcertoTM (70)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00847536033849), received between Jul 16, 2024 and Jul 16, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Separation Failure” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 16, 2024Separation Failure; Separation Failure
MEDTRONIC RECEIVED REPORT THAT TWO CONCERTO COIL COULD NOT BE DETACHED. IT WAS NOTED THAT THE DEVICES WERE PREPARED PER THE INSTRUCTIONS FOR USE (IFU). THE COILS WERE NOT IMPLANTED AND WERE REMOVED AND REPLACED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WAS NO PATIENT INJURY. ANCILLARY DEVICE: PROGREAT 2.4 MICROCATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19748925 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page