ConcertoTM PV-18-40-3D (UDI 00847536033863)Recalls & reported problems
Micro Therapeutics, Inc. · COIL PV-18-40-3D V04 · Class 2 · Implant
All models of ConcertoTM (70)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00847536033863), received between Jan 10, 2021 and May 6, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Separation Failure” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Separation Failure3
- Stretched1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 6, 2024Stretched; Separation Failure; Stretched; Separation Failure
ADDITIONAL INFORMATION WAS RECEIVED. AT LEAST 3 ATTEMPTS WERE MADE TO SEPARATE THE COIL USING THE INSTANT DETACHER. 2 MORE ATTEMPTS WERE MADE TO SEPARATE THE COIL USING THE MANUAL METHOD. PRIOR TO THE NON-REMOVAL OF THE COIL, THERE WERE NO PROBLEM. THE PUSHER WIRE WAS OBSERVED TO BE STRETCHED AFTER REMOVAL FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19256447 ↗MalfunctionMar 25, 2021Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THE CONCERTO COIL COULD NOT BE DETACHED WITH THE INSTANT DETACHER (ID) OR THE MANUAL BREAK METHOD. THE PATIENT WAS UNDERGOING TREATMENT FOR INTERNAL BLEEDING. THE PATIENT'S BLOOD FLOW AND VESSEL TORTUOSITY WERE NORMAL. IT WAS REPORTED THAT THE TWO DETACHMENT ATTEMPTS WERE MADE WITH THE ID, AND ONE ATTEMPT WAS MADE MANUALLY. THERE WAS SAID TO BE NO ISSUE WITH THE ID, AND ANOTHER ID WAS NOT USED. THE COIL WAS REPLACED, AND THE PATIENT DID NOT EXPERIENCE ANY INJURY OR COMPLICATIONS. THE DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11572930 ↗MalfunctionJan 10, 2021Separation Failure; Separation Failure
MEDTRONIC RECEIVED A REPORT THAT THE 18X40 BECAME STUCK AND BUNCHED/DAMAGED IN THE MIDDLE OF THE MICROCATHETER. IT WAS NOTED THAT TH E CATHETER WAS DAMAGED IN THE PROXIMAL END. A SECOND COIL 20X50 WAS USED WHICH WOULD NOT DETACH MANUALLY. 3 DETACHMENT ATTEMPTS WERE MADE WITH THE INSTANT DETACHER, AND ONLY ONE INSTANT DETACHER WAS USED. THE PATIENT WAS UNDERGOING TREATMENT OF AN UNRUPTURED, SACCULAR SPLENIC ANEURYSM WITH A MAX DIAMETER OF 25MM. THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MODERATE. THE DEVICES WERE PREPARED AS INDICATED PER THE IFU. THERE WERE NO RELATED PATIENT SYMPTOMS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11144581 ↗
Other Device, Vascular, for Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.