ConcertoTM PV-18-40-3D (UDI 00847536033863)Recalls & reported problems

Micro Therapeutics, Inc. · COIL PV-18-40-3D V04 · Class 2 · Implant

All models of ConcertoTM (70)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00847536033863), received between Jan 10, 2021 and May 6, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Separation Failure” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Separation Failure3
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 6, 2024Stretched; Separation Failure; Stretched; Separation Failure

    ADDITIONAL INFORMATION WAS RECEIVED. AT LEAST 3 ATTEMPTS WERE MADE TO SEPARATE THE COIL USING THE INSTANT DETACHER. 2 MORE ATTEMPTS WERE MADE TO SEPARATE THE COIL USING THE MANUAL METHOD. PRIOR TO THE NON-REMOVAL OF THE COIL, THERE WERE NO PROBLEM. THE PUSHER WIRE WAS OBSERVED TO BE STRETCHED AFTER REMOVAL FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19256447 ↗
  • MalfunctionMar 25, 2021Separation Failure; Separation Failure

    MEDTRONIC RECEIVED A REPORT THAT THE CONCERTO COIL COULD NOT BE DETACHED WITH THE INSTANT DETACHER (ID) OR THE MANUAL BREAK METHOD. THE PATIENT WAS UNDERGOING TREATMENT FOR INTERNAL BLEEDING. THE PATIENT'S BLOOD FLOW AND VESSEL TORTUOSITY WERE NORMAL. IT WAS REPORTED THAT THE TWO DETACHMENT ATTEMPTS WERE MADE WITH THE ID, AND ONE ATTEMPT WAS MADE MANUALLY. THERE WAS SAID TO BE NO ISSUE WITH THE ID, AND ANOTHER ID WAS NOT USED. THE COIL WAS REPLACED, AND THE PATIENT DID NOT EXPERIENCE ANY INJURY OR COMPLICATIONS. THE DEVICES WERE PREPARED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11572930 ↗
  • MalfunctionJan 10, 2021Separation Failure; Separation Failure

    MEDTRONIC RECEIVED A REPORT THAT THE 18X40 BECAME STUCK AND BUNCHED/DAMAGED IN THE MIDDLE OF THE MICROCATHETER. IT WAS NOTED THAT TH E CATHETER WAS DAMAGED IN THE PROXIMAL END. A SECOND COIL 20X50 WAS USED WHICH WOULD NOT DETACH MANUALLY. 3 DETACHMENT ATTEMPTS WERE MADE WITH THE INSTANT DETACHER, AND ONLY ONE INSTANT DETACHER WAS USED. THE PATIENT WAS UNDERGOING TREATMENT OF AN UNRUPTURED, SACCULAR SPLENIC ANEURYSM WITH A MAX DIAMETER OF 25MM. THE PATIENT'S BLOOD FLOW WAS NORMAL AND VESSEL TORTUOSITY WAS MODERATE. THE DEVICES WERE PREPARED AS INDICATED PER THE IFU. THERE WERE NO RELATED PATIENT SYMPTOMS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11144581 ↗